8 Guideline value for power settings
Monopolar cutting:
Recommended settings
Smooth cuts with thin needle electrode
5 - 40 W
Cutting with knife electrode, lancet or wire
loop
50 - 175 W
Cutting in wet environment (TUR)
80 - 150 W
Monopolar coagulation:
Recommended settings
Coagulation of large surfaces
60 W
Coagulation of small surfaces
40 W
Bipolar application:
Recommended settings
Coagulation
10 - 30 W
Cutting
40 - 60 W
Cutting in water environment (TUR)
50 - 75 W
When performing the power settings, the user should generally
take into account the cutting speed, geometry of the electrode
and properties of the tissue.
!
The recommended settings serve as guideline values only and
must be adjusted to the relevant situation.
In the event of unfavourable conditions, these values may
deviate!