HBS x-touch 300
45
EN
7.3 Foot switch settings
After connecting the foot switch at the back, the required mode must be set at the touchscreen. To set
the mode, touch the field indicating the foot switch symbol. This changes the text which appears
above the symbol.
It is important to ensure that the instruments are connected to the connectors, which are selected by
footswitch. For example: In Monopolar 1 connected instruments can only be operated with the
settings MONO1/MONO2 or MONO1 + BIAUTO of the foot switch.
7.4 Handles with buttons
Activating the handle with buttons takes place independently of the foot switch settings. These cannot be
activated using the foot switch.
7.5 Neutral electrode settings and attachment
In the monopolar operating mode, a neutral electrode is required.
After attaching the neutral electrode to the patient, the HF unit recognize the neutral
electrode automatically (cf. illustration on the left and chapter 4.2
”).
The HBS x-touch 300 recognizes neutral electrodes with
divided and undivided contact surfaces.
!
If the displayed neutral electrode type does not agree with the used
neutral electrode type, this is indicative of poor adhesion of the
neutral electrode on the patient, a defective neutral electrode or a
defective cable!
Please note the following information when attaching the
neutral electrode
Summary of Contents for HBS 100
Page 2: ...HBS x touch 300 2 DE ...
Page 4: ...HBS x touch 300 4 DE ...
Page 30: ...HBS x touch 300 30 DE ...
Page 32: ...HBS x touch 300 32 EN ...
Page 58: ...HBS x touch 300 58 EN ...
Page 60: ...HBS x touch 300 60 FR ...
Page 86: ...HBS x touch 300 86 FR ...
Page 113: ...HBS x touch 300 113 ES ...
Page 115: ...HBS x touch 300 115 IT ...