The following error messages can occur:
If one or more numerical codes are displayed, or if the device is indicating any other recognizable
defects, please refer to [email protected].
◼
HF current does not flow when the foot switch is actuated. The system only reacts with
a tone signal and the display changes colour.
Please check:
➔
That all terminals have been correctly connected
➔
The settings of the foot switch
➔
Whether the neutral electrode has been recognized
Error
No. Cause of error
Remedying errors
Err 1
Total Failure
1
"Power timeout error"
The supply voltage is not achieved.
Check the controller
and the power supply unit, exchange if applicable
10
"Power Error"
The output measured by the monitoring module is 20%
greater than the set output. Check the controller, the
interface and the power supply unit, exchange if applicable.
Err 2
Modules
Failure
2
"NE electrode error"
Check the terminal, and the plate on the patient,
where applicable exchange NE and/or connecting
cable.
6
Mono - Cut hand /FS1
Error”
Check the CUT terminals, check the settings,
exchange accessories if necessary. If error persists,
inform specialist personnel.
7
"Mono
– Coag (Bi) hand
/FS1 Error"
Check the COAG terminals, check the settings,
exchange accessories if necessary. If error persists,
inform specialist personnel.
Err 03
PS Failure
4
"Power Supply 1 Error"
The necessary voltage from the block for regulating the
power voltage for the output generator in idle running is
missing. Check the power terminal and power supply unit,
exchange if applicable.
5
"Power Supply 2 Error"
The necessary voltage from the block is missing for
correction of the output factor. Check the power terminal
and power supply unit, exchange if applicable.
Err 5
CPU Failure
11
"EEROR in CPU ROM"
Incorrect checksum of the ROM memory from control
processor. Check controller, exchange if applicable
12
"EEROR in CPU RAM"
Incorrect checksum of the RAM memory from control
processor. Check controller, exchange if applicable.
Err 6
Keys Failure
0-
17
„Button stick“
Check your settings and the buttons.