Airborne 360°-300 / Airborne 360°-300UV Air purification unit
35
EN
9.3.2 Changing the H14 filter
1. Remove the service cover by loosening the 2 screws on the sides with the help of an Allen
key.
2. Remove the front cover of the unit by loosening the 6 screws on the sides with the help of an
Allen key.
3. Remove the H14 filter from the unit. Dispose of it, in an environmentally friendly manner, in the
foil bag of the new filter.
4. Insert new H14 filter.
5. Attach the front cover, then attach the service cover.
9.3.3 Changing UVC lamps
NOTE! When replacing the UVC lamps, make sure that they are only touched with gloves.
Otherwise, the lifetime of the lamps may be adversely affected.
Operating time of UVC lamps: 9000h.
1. Remove the front cover of the F7 filter by loosening both screws with the help of an Allen key.
2. Remove the front cover of the unit by loosening all 6 screws on the sides with the help of an
Allen key.
3. Remove UVC lamp. Dispose of it in an environmentally friendly manner.
NOTE! The lamps must always be replaced in pairs, even if only one lamp is defective.
4. Insert a new UVC lamp. It must audibly engage.
5. Attach the front cover, then attach the service cover.