Troubleshooting

15

If any abnormalty occurs during use, please check and correct the following:

Condition

Correction

Display is blank when power is on.

Check and correct the polarity of the installed
batteries. Reinstall or replace batteries.

Measurement incomplete or

Review and follow “Applying Your Blood

abnormally low or high values

Pressure Cuff” and “Taking Your Blood Pressure

displayed.

Reading” sections. Replace the batteries.

EE symbol displays.

Do not talk or move during the measurement.
Review “Applying Your Blood Pressure
Cuff” section.

Measurements are different from

Remember, blood pressure readings are

those typically measured by

influenced by physical and mental conditions

physician or every measurement

and/or even the time of day. Daily

reading is different.

measurements should be interpreted by your
physician.

Indicates a measurement
within Stage 3 Hypertension.

Indicates a measurement
within Stage 2 Hypertension.

Indicates a measurement
within Stage 1 Hypertension.

Appears when batteries should be
replaced.

Shows the pulse rate per minute.

Occurs when an error was made
during measurement.

Appears when there is an irregular
heartbeat detected during
measurement.

Appears when the unit is ready or a
measurement error has occurred. Restart
the measurement if an error occurs.

Appears when the cuff is unstable or it
has too much air left in it. Once the arrow
stops flashing, the unit is ready to take a
measurement.

Appears when recalling a
measurement from memory.

DIA

SYS

DIA

SYS

DIA

SYS

Description of Display Symbols

14

You can recall up to 120 total measurements in memory (40 per
memory bank). These measurements can be shared with your physician
or trained healthcare professional.

1. Press and release the MEMORY button. The unit will display U1, U2 or

U3 depending upon the user bank selected. It will also show how
many measurements are stored within that bank.

2. To select the appropriate user bank press the START button to switch

between U1, U2 or U3.

3. Once the desired user memory bank is selected, press the MEMORY

button again to scroll through the saved measurements,

Fig 1

. After

you’ve scrolled through all the readings, the unit will automatically
turn off.

NOTE:

Each memory bank can store up to 40 readings. When the

number of readings exceeds 40, the oldest data will be replaced with
the new record.

To erase all stored measurements:

1. Press and release the MEMORY button. The unit will display U1, U2 or

U3 depending upon the user bank selected and how many
measurements are stored within that bank.

2. Select the desired user bank by pressing and releasing the MEMORY

button.

3. Press and hold down the Memory button until two horizontal lines

flash,

Fig 2

. This indicates that all measurements within the selected

memory bank have been erased.

Fig. 1

Fig. 2

Recalling Measurements in Memory

Summary of Contents for 04-620-001

Page 1: ...Women s Automatic Digital Blood Pressure Monitor Instruction Manual Please read this guidebook completely before operating this unit Item 04 625 001 English Spanish ...

Page 2: ...warranty stated above The warrantor will not be responsible for any indirect incidental special consequential or punitive damages or other loss including but not limited to damage to or loss of other property or equipment and personal injuries whether to purchaser or others The warrantor shall in no event be liable to the purchaser for any amount in excess of the cost of repair and or replacement ...

Page 3: ...display both systolic and diastolic readings on the screen Your systolic will be positioned as the upper number and the diastolic reading will be the lower number systolic 120 80 diastolic Blood pressure is measured in millimeters mm of mercury Hg and is generally recorded with the systolic pressure 120 listed first and the diastolic pressure 80 listed second The numbers are typically separated by...

Page 4: ...ed by pressing the Release button 7 Why Monitor Your Blood Pressure at Home A visit to a physician s office can be a stressful situation for a patient And anxiety is a known factor in raising blood pressure This temporary occurrence of elevated blood pressure at the physician s office is commonly referred to as white coat syndrome Whether or not you experience white coat syndrome home blood pressu...

Page 5: ...ressure 85 89 mmHg Normal Systolic pressure 120 129 mmHg or Diastolic pressure 80 84 mmHg Optimal Systolic pressure 120 mmHg and Diastolic pressure 80 mmHg DIA SYS DIA SYS DIA SYS DIA SYS DIA SYS DIA SYS World Health Organization WHO Indicator 9 Irregular Heartbeat Detection Your digital blood pressure monitor features Irregular Heartbeat Detection This feature allows users to accurately monitor b...

Page 6: ...idnight 0 12 pm Noon 12 1 am 1 1 pm 13 2 am 2 2 pm 14 3 am 3 3 pm 15 4 am 4 4 pm 16 5 am 5 5 pm 17 6 am 6 6 pm 18 7 am 7 7 pm 19 8 am 8 8 pm 20 9 am 9 9 pm 21 10 am 10 10 pm 22 11 am 11 11 pm 23 Setting the Date Time 10 Battery Installation Replacement 1 Press down on tab and lift the cover 2 Insert or replace 4 AA batteries into the compartment matching the indicated polarity symbols Fig 1 3 To c...

Page 7: ...cally reinflate to approximately 40 mmHg higher than the last pressure level 13 Fig 3 Fig 6 Fig 5 Fig 4 Fig 1 Fig 2 Fig 7 Taking Your Blood Pressure Reading Before applying your blood pressure cuff be sure you have selected the appropriate cuff Standard Cuff fits arm circumference 11 3 4 16 1 2 30 42 cm Large Cuff fits arm circumference 16 1 2 18 7 8 42 48 cm NOTE If for any reason you are unable ...

Page 8: ...the cuff is unstable or it has too much air left in it Once the arrow stops flashing the unit is ready to take a measurement Appears when recalling a measurement from memory DIA SYS DIA SYS DIA SYS Description of Display Symbols 14 You can recall up to 120 total measurements in memory 40 per memory bank These measurements can be shared with your physician or trained healthcare professional 1 Press...

Page 9: ... Measuring range Pressure 0 300 mmHg Pulse 40 180 beats minute Accuracy Pressure 3 mmHg Pulse 5 of reading Inflation Automatic Inflation Deflation Automatic Pressure Release Valve Manual fast release Memory Built in memory enabling display of up to 120 measurements Automatic shut off Approximately 1 minute after last button operation Operation environment Temperature 41o F 104o F 5o C 40o C Humidi...

Page 10: ...Manual de Instrucciones Monitor Digital de Presión Arterial Automático Mujer Sírvase leer toda esta guía antes de operar la unidad Artículo n º 04 625 001 Inglés Español Blood Pressure Log 18 ...

Page 11: ...proveer con o usar con el Monitor Digital de Presión Arterial Esta garantía no tendrá validez si el producto Monitor Digital de Presión Arterial no se usa correctamente o se maltrata de cualquier manera Si el Monitor Digital de Presión Arterial presenta fallas de operación durante los primeros cinco años a partir del momento original de compra llame a nuestra Línea de Ayuda de Atención al Cliente ...

Page 12: ...a presión diastólica sistólica 120 80 diastólica La presión arterial se mide en milímetros mm de mercurio Hg y por lo general al registrarla se indica primero la presión sistólica 120 y la diastólica 80 después Típicamente los números están separados por una barra según se indica más arriba Ambas lecturas la SISTóLICA y la DIASTóLICA son necesarias para que un médico pueda evaluar el estado de la ...

Page 13: ... inflado y desinflado con sólo pulsar el botón de ENCENDIDO APAGADO 24 Por Qué Controlar la Presión Arterial en Hogar La visita al consultorio de un médico puede resultar estresante para un paciente Y la ansiedad es un factor que se sabe eleva la presión arterial Esta situación temporal de presión arterial elevada en e consultorio del médico se conoce comúnmente con el nombre de síndrome del delan...

Page 14: ... de Hipertensión Presión arterial sistólica 160 179 mmHg o Presión arterial diastólica 100 109 mmHg Grado 1 de Hipertensión Presión arterial sistólica 140 159 mmHg o Presión arterial diastólica 90 99 mmHg Alta Normal Presión arterial sistólica 130 139 mmHg o Presión arterial diastólica 85 89 mmHg Normal Presión arterial sistólica 120 129 mmHg o Presión arterial diastólica 80 84 mmHg Óptima Presión...

Page 15: ...r 12 horas Hora formato militar 24 horas Hora formato estándar 12 horas Hora formato militar 24 horas 12 am Medianoche 0 12 pm Mediodía 12 1 am 1 1 pm 13 2 am 2 2 pm 14 3 am 3 3 pm 15 4 am 4 4 pm 16 5 am 5 5 pm 17 6 am 6 6 pm 18 7 am 7 7 pm 19 8 am 8 8 pm 20 9 am 9 9 pm 21 10 am 10 10 pm 22 11 am 11 11 pm 23 28 1 Presione sobre la lengüeta y levante la cubierta 2 Coloque o cambie 4 baterías AA en ...

Page 16: ...l de presión anterior 31 Cómo Tomar la Lectura de la Presión Arterial Fig 1 Fig 2 Antes de colocarse el brazalete para la presión arterial verifique que tenga el brazalete apropiado Brazalete para adulto corresponde a circunferencia de brazo 11 3 4 16 1 2 30 42 cm Brazalete para adulto grande corresponde a circunferencia de brazo 16 1 2 18 7 8 42 48 cm NOTA Si por algún motivo no puede o no debe u...

Page 17: ...a flecha ha dejado de titilar la unidad está lista para tomar una medición Aparece cuando trae una medición de la memoria DIA SYS DIA SYS DIA SYS Descripción de los Símbolos del Visor Análisis de Fallas 32 Cómo Llamar Mediciones de Memoria Puede recuperar un total de hasta 120 mediciones en la memoria 40 por banco de memoria Estas mediciones pueden compartirse con su médico o con un profesional ca...

Page 18: ... de Medición Oscilométrico Fuente de energía 4 baterías tipo AA Alcance de la medición Presión 0 300 mmHg Pulso 40 180 pulsaciones minuto Precisión Presión 3 mmHg Pulso 5 de la lectura Inflado Inflable automático Memoria Visor con memoria incorporada que permite hasta 120 mediciones Apagado automático Alrededor de un minuto después de presionar el último botón Ambiente de funcionamiento Temperatur...

Page 19: ...Registro de Presión Arterial 36 ...

Reviews: