drying.
Maintenance, inspection, and testing:
Assembly of Spray Gun
1.
Lubricate the front end of the piston and the small piston O-ring with pure silicone oil
prior to the assembly. Only minimal lubrication! If you insert the dry piston the piston
seal will be damaged!
2.
Then insert the piston into the gun bod, position the piston spring onto the piston
and screw the end cap to the gun body.
3.
Screw the nozzle cap to the front of the gun body and screw the desired nozzle into the
nozzle cap.
4.
Screw the gun handle into the gun body as far as it will go.
5.
Oil the plug nipple of the handle DN7.2 with pure silicone oil if it is difficult to insert
it into or remove it from stainless steel coupling.
Disassembling the cleaning gun
:
1.
Disassemble the end cap from the body of the gun by turning it counterclockwise.
2.
Remove the piston spring.
3.
Pull trigger back and pull the piston out backwards.
4.
Loosen nozzles from nozzle cap by turning them counterclockwise.
5.
Disassemble the nozzle cap from the body of the gun by turning it counterclockwise.
6.
Disassemble the handle from the body of the gun by turning it counterclockwise.
7.
Disassemble the shower nozzle sieve from the shower nozzle body by turning it
counterclockwise.
Reassembly of the cleaning gun to the gun coupling:
1.
Introduce the gun handle with coupling plug DN7.2 straight into the gun coupling,
and press it in until it audibly engages into position.
2.
Open the pressure supply line and check the whole system for leakage!
Dismounting the cleaning gun from the gun coupling:
1.
If you want to dismount the cleaning gun from the coupling again for reprocessing,
you must interrupt the pressure supply.
2.
Relieve the pressure in the hose through the gun by activating the trigger.
3.
Press the gun handle into the coupling and, at the same time, pull the sliding ring and
top of coupling down coupling towards the hose and pull the gun out of the coupling.
4.
Put the sterile coupling protection cap over the gun coupling.
5.
The hose must be stored in such a way that it does not come into contact with liquids
and cannot be contaminated.
6.
The coupling protection cap is cleaned and sterilized in the same way as the cleaning
gun parts.
Maintenance
1.
It is recommended to decalcify the Spray Gun approximately every four weeks. The
gun should be disassembled and the inner parts placed in a decalcifying agent. This is
beneficial for the service life and reliability of the spray gun.
2.
If it is difficult to remove the cg piston from the gun body or insert it into the gun body
then you must replace the small O-ring for the cg piston.
3.
Replacement parts and instructions for repair are available from Healthmark.
Reassembly Instructions:
Packaging:
Sterilization:
Storage:
Additional Information:
Caution: Oils containing white oil or paraffin oil will damage the O-rings.