background image

 

 

 

 

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

 

 
 
 
 

MODEL:  594KL 

 

 

Thank you for your purchase of this product. 

Please read this manual carefully and keep it handy for ready reference.  

 

 

TABLE OF CONTENTS

 

 

CAUTION AND WARNING ........................................................ 2 
SCALE SPECIFICATIONS ........................................................ 2 
SET UP ....................................................................................... 3 
OPERATING INSTRUCTIONS .................................................. 3 
PRESERVING BATTERY/PROGRAMMING AUTO-OFF TIME.. 3 
MAINTENANCE.......................................................................... 3 
TROUBLESHOOTING................................................................ 4 
CALIBRATION PROCEDURE.................................................... 4 
LIMITED WARRANTY................................................................ 5 

 
 

  

To prevent injury and damage to your scale, please follow these instructions very carefully

 

Do not transport the scale while the chair is loaded. 
For accurate weighing the scale must be placed on a flat, stable surface. 
For accurate weighing, verify proper operation according to the procedure described in this manual before each use. 
Do not use in the presence of flammable materials. 
Operating at other voltages and frequencies than specified could damage the equipment. 
Do not leave patient unattended in the chair. 
Do not use chair scale in place of a wheelchair to transport patients. 
Do not bounce scale down stairs, as doing so may damage internal parts. 
Do not use the footrest as a step. 
To maximize battery life, do not leave charger plugged into the scale for extended periods.  

 

GENERAL

 

 
Health o meter Pr

ofessional’s Digital Chair Scale, model 594KL, uses sophisticated microprocessor technology. This precision instrument is 

designed to provide accurate, reliable and repeatable weight measurements and features that make the weighing process simple, fast and 
convenient. The scale is set up to determine the weight of a stationary patient. The weight can be displayed in pounds or in kilograms. The 
unit can be operated using its 6 Volt rechargeable battery. 

 

SCALE SPECIFICATIONS

 

 

Capacity and resolution

 

600 lb x 0.2 lb / 270 kg x 0.2 kg

 

Power Requirements

 

6 Volt Rechargeable battery, charger and battery included

 

Environmental

 

Operating temperatures: 5°C to 40°C (41°F to 104°F) 
Storage temperature: -1°C to +52°C (30°F to 125°F) 
Humidity: 85% rH

 

Chair

 

18 1/4

” (w) x 15” (d) x 17 ½” (h)

 

Scale Dimensions

 

24 3/8” (w) x 32 ½” (d) x 36 ¾” (h) (Arm Width 20 1/2”)

 

CAUTION AND WARNING 

SCALE SPECIFICATIONS 

Summary of Contents for 594KL

Page 1: ...TING INSTRUCTIONS DIGITAL CHAIR SCALE MODEL 594KL Please register your product at www homscales com PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2016 P N UM594KL Rev 20160308 ...

Page 2: ...eelchair to transport patients Do not bounce scale down stairs as doing so may damage internal parts Do not use the footrest as a step To maximize battery life do not leave charger plugged into the scale for extended periods GENERAL Health o meter Professional s Digital Chair Scale model 594KL uses sophisticated microprocessor technology This precision instrument is designed to provide accurate re...

Page 3: ...or 3 seconds toggles between kilograms or pounds 4 TARE ZERO Zeros the scale prior to weighing or sets a tare value 5 ON OFF Turns scale ON and OFF The 594KL is equipped with a power switch located on the side of the display When the switch is moved to the On I position it is supplying power to the scale even when the display head is off To charge the battery this switch must be in the On I positi...

Page 4: ...ghts The recommended calibration weight is 600 lbs however calibration can be performed using any weight no less than 44 lbs and no more than 600 lbs 1 Turn the scale on 2 At the same time press the TARE ZERO and KG LB buttons t and the number representing the current automatic shut off value will appear on the display 3 Press the TARE ZERO button CAL and then 600 0 will appear on the display 4 To...

Page 5: ...ct use on improper voltage or current use contrary to the operating instruction abuse including tampering damage in transit or unauthorized repair or alternations Further the warranty does not cover natural disasters such as fire floods hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province...