Health O Meter 549KG User Instructions Download Page 5

P/N UM549KL_549KG Rev20190109

 

OPERATING INSTRUCTIONS 

Weighing a Patient  

1.  Make sure there is nothing on the weighing tray. Press the 

ON/TARE/OFF

 button to turn on 

the scale

. “88.88” will appear on the display followed by zeroes.  

2.  Place the baby on the scale so that its weight is evenly distributed. The scale will begin to 

display weight information and will lock on the patient’s weight when the stable weight of the 
patient is determined.  Depending on the movement of the patient on the scale, it may take 
several seconds for the scale to lock onto the weight.  When a stable weight is calculated, 
the weight reading will flash on the display three times. The weight reading will remain on 
the display until the baby is removed from the scale. 

Caution: To prevent patient injury, 

the patient must be attended throughout the entire weighing event.   

3.  Remove the baby from the scale. 
4.  To turn off the scale, press and hold the 

ON/TARE/OFF

 button for approximately three 

seconds.  

Note:

 I

f the display shows “Err” you have exceeded the capacity of the scale. 

 
Tare Function  

The tare function allows you to place other items on the scale while weighing the baby without 

affecting the baby’s weight reading. 
1.  Make sure there is nothing on the weighing tray. Press the 

ON/TARE/OFF

 button to turn on 

the scale. “88.88” will appear on the display followed by zeroes.  

2.  Place the object you want to tare on the scale. 

The display will show a value for the object’s 

weight. 

3.  With the object remaining on the scale, press the 

ON/TARE/OFF

 button.  

“TARE” will 

appear on the display then the scale will return to zero. 

4.  Keep the object on the scale and place the baby on the scale. The scale will automatically 

deduct the weight of the object and only 

display the baby’s weight. 

Note:

 If the scale does not return to zero when you remove the object and baby from the tray, 

press the 

ON/TARE/OFF

 button to return the scale to zero. 

 

LB/KG Lock (549KL ONLY) 

The weight measuring unit can be locked to only measure in LB or KG. 
 

KG Lock Instructions 

1.  Make sure there is nothing on the weighing tray. Press the 

ON/TARE/OFF

 button to turn on 

the scale. “88.88” will appear on the display followed by zeroes.  

2.  Press the 

kg/lb

 button so the KG icon is displayed and continue holding the 

kg/lb

 button for 

approximately 10 seconds. 

3.  When the KG icon flashes, release the 

kg/lb

 button. The KG weight measuring unit is now 

locked and will remain locked even if the scale is turned off. 

 

LB Lock Instructions 

1.  Make sure there is nothing on the weighing tray. Press the 

ON/TARE/OFF

 button to turn on 

the scale. “88.88” will appear on the display followed by zeroes.  

2.  Press the 

kg/lb

 button so the LB icon is displayed and continue holding the 

kg/lb

 button for 

approximately 10 seconds. 

3.  When the LB icon flashes, release the 

kg/lb

 button. The LB weight measuring unit is now 

locked and will remain locked even if the scale is turned off. 
 

Disabling LB/KG Lock  

1.  Make sure there is nothing on the weighing tray. Press the 

ON/TARE/OFF

 button to turn on 

the scale. “88.88” will appear on the display followed by zeroes.  

2.  Press and hold the 

kg/lb

 button until LB and KG icons 

both

 flash on the display. When both 

LB and KG icons are flashing, release the 

kg/lb

 button. The weight measuring unit is now 

unlocked; pressing the 

kg/lb

 button will allow user to toggle between LB and KG.   

Summary of Contents for 549KG

Page 1: ...P N UM549KL_549KG Rev20190109 1 Model 549KL 549KG Digital Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2019...

Page 2: ...s 3 Specifications 4 Operating Instructions 5 Battery Installation Maintenance 6 Calibration 549KL 7 Calibration 549KG 8 Warranty 9 Note This scale has been factory calibrated and does not require cal...

Page 3: ...ch use the proper operation according to the procedure described in this manual Do not transport the scale with a patient or object on the tray Do not exceed the weight capacity specified for this sca...

Page 4: ...o 50 C Atmospheric Pressure 500hPa 1060hPa Humidity 10 95 RH Physical Dimensions Tray Size Width 19 1 4 489 mm Depth 13 330 mm Height 2 5 8 67 mm Product Footprint Width 19 3 8 492 mm Depth 13 1 2 343...

Page 5: ...the scale will return to zero 4 Keep the object on the scale and place the baby on the scale The scale will automatically deduct the weight of the object and only display the baby s weight Note If the...

Page 6: ...l scale for any obvious damage wear and tear If the scale does not operate correctly refer to qualified service personnel Note This scale is an extremely sensitive weighing device Items such as clothi...

Page 7: ...bers When the calibration numbers stabilize press the ON TARE OFF button C10 will flash on the display C10 Place 10 pounds on the scale After stabilization C20 will flash on the display C20 Place an a...

Page 8: ...scale off press and hold the ON TARE OFF button and the CAL button at the same time When CAL appears on the display release both buttons Numbers will appear indicating you can begin calibration CAL f...

Page 9: ...od hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to ju...