13
CALIBRATION
This scale has been factory calibrated, and does not require calibration prior to
use.
If required, the scale can be calibrated. Calibration of your scale is performed
using kilograms (KG). 20 kilograms are required for calibration, applied in 5 kg and
10 kg increments.
Note: Only weights certified and traceable to national standards should be used
for calibration procedures.
ACTION
DISPLAY
SHOWS
With the scale off, press and hold the
LB/KG
button (524KL) or the
blank blue button (524KG) and
ZERO
buttons at the same time, then
press the
ON/OFF
button to turn on the scale. “CAL” will briefly
appear on the display, followed by four numbers indicating that you
can begin calibration. Release all three buttons.
“CAL” followed by
“XXXX”
Press the
ZERO
button. “C-5” will flash on the display, followed by
“0”or “1”.
“C-5” then “0” or
“1”.
Place 5 kg on the scale. Four numbers will appear. When calibration
numbers stabilize, press the
ZERO
button. “C-15” will appear on the
display,
followed by “0” or “1”.
“XXXX” followed
by
“C-15” then “0”
or “1”.
Place an additional 10 kg on the scale, for a total of 15 kg. Four
numbers will appear. When calibration numbers stabilize, press the
ZERO
button. “C-20” will appear on the display, followed by “0” or “1”.
“XXXX” followed
by
“C-20” then “0”
or “1”.
Place an additional 5 kg on the scale, for a total of 20 kg. Four
numbers will appear. When calibration numbers stabilize, press the
ZERO
button.
“END” will appear on the display, followed by “OFF”.
The scale will automatically shut off.
“XXXX” followed
by “END” then
“OFF”