background image

 

11 

OPERATING INSTRUCTIONS

 

WEIGHING A PATIENT 

– MODELS 524KL AND 524KG 

1.  Make sure there is nothing on the weighing seat. 
2.  Press the 

ON/OFF

 

button to turn the display assembly on. “88:8:8.8” will appear on the 

display followed by zeroes.  

3.  Place the baby on the weighing seat so that its weight is evenly distributed. The scale will 

begin to display weight information and will lock on the patient’s weight when the stable 
weight of the patient is determined.  Depending on the movement of the patient on the scale, 
it may take several seconds for the scale to lock onto the weight.  When a stable weight is 
calculated, 

a small ring will appear on the display along with the baby’s weight. If the display 

shows “OVER” you have exceeded the capacity of the scale. 

Caution: To prevent patient injury, the patient must be attended throughout the entire 
weighing event.   

4.  The weight may be held on the display by pressing the 

HOLD/RELEASE

 button.  To 

release, press the 

HOLD/RELEASE 

button again. 

5.  After the weight measurement has been taken and is recorded, remove the baby from the 

scale. 

6.  Press the 

ON/OFF

 button to turn the scale off. 

 
ZERO FUNCTION 

– MODELS 524KL AND 524KG 

Zeroes the scale prior to a weight measurement. The 

ZERO/TARE

 button is also used in the 

Tare function as outlined below.  
 

TARE FUNCTION 

– MODELS 524KL AND 524KG 

The tare function allows you to place other items on the scale while weighing the baby without 
affecting the baby’s weight reading. 
1.  With all weight off the scale, press the 

ON/OFF

 button to turn on the scale. 

2.  Wait until zeroes appear on the display. 
3. 

Place the object on the scale. The display will show a value for the object’s weight. 

4.  Press the 

ZERO/TARE

 

button.  The word “TARE” will appear on the display, followed by 

zeroes. 

5.  Keep the object on the platform, and place the baby on the weighing seat. The scale will 

display the baby’s weight. 

Note: If the scale does not return to zero when you remove the object and baby from the 
weighing seat, press the 

ZERO/TARE

 button to return the scale to zero. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for 524KG

Page 1: ...1 Model 524KL 524KG Digital Pediatric Seat Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM524KL_524KG Rev20180227...

Page 2: ...l 5 Assembly Instructions 6 Powering the Scale 9 Set Up Instructions 10 Operating Instructions 11 Maintenance 12 Calibration 13 Troubleshooting 14 Warranty 15 Note This scale has been factory calibrat...

Page 3: ...e of an unspecified adapter will void the warranty and can pose a serious safety hazard Prior to using this scale inspect the AC adapter cord for cracking fraying or for broken bent plug prongs Prior...

Page 4: ...0 20 lb 0 2 oz 20 50 lb 0 5 oz 0 9 kg 5g 9 23 kg 10 g 524KG 23 kg 0 9 kg 5g 9 23 kg 10 g Power Requirements 6 AA batteries included optional 120V power adapter ADPT31 Environmental Operating temperat...

Page 5: ...tivity Solutions This connectivity helps minimize transcription errors when recording patient data For further information regarding how to connect this scale to other electronic devices please contac...

Page 6: ...ind inside the main carton as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton asid...

Page 7: ...column and pull gently until the plug reaches the bottom of the column Carefully pull the cable connector from the bottom of the column and plug into the RJ jack on the scale platform Caution Once con...

Page 8: ...brackets and into the platform Using a Phillips head screwdriver tighten the screws to secure both brackets to the platform IMPORTANT Ensure that bracket C is mounted on the edge of the platform not t...

Page 9: ...Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a problem with the weight measu...

Page 10: ...te in the selected mode 5 Repeat steps 3 4 to change the weighing mode to a different unit of measure LB KG Lock Out The weight measuring unit can be locked to only measure in KG or LB KG Lock Instruc...

Page 11: ...e HOLD RELEASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the scale 6 Press the ON OFF button to turn the...

Page 12: ...r to prevent battery depletion do not store or leave anything on the scale Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation Caution Disconnect t...

Page 13: ...ating that you can begin calibration Release all three buttons CAL followed by XXXX Press the ZERO button C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four...

Page 14: ...eighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not sitting still Try to have the patient sit still The display shows Over messa...

Page 15: ...d tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To ge...

Reviews: