background image

P/N UM402LB REV20190501 

11 

TRANSPORTATION

 

 
A wheel bracket is included if purchased as model 402LBWH. If purchased as model 402LB an 
optional wheel bracket can be purchased separately (Item #55000). Follow these steps to install 
the wheel bracket: 

 

Tools Required (not included) 

Adjustable wrench or 10 mm wrench 
 
1. Remove the mounting screws that attach the 

clamping plate to the wheel bracket. 

 
 
 
 
 
 
 
 
 
 
 

2. Gain access to the underside of the scale 

base by carefully tipping the scale forward 
until the pillar is resting on the floor. Position 
the clamping plate inside the base by 
aligning the holes in the clamping plate and 
base.  

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

3. Position the wheel bracket onto the base by 

aligning the holes in the wheel bracket with 
the holes in the clamping plate. 

 
 
 
 
 
 
 

4. Insert the mounting screws through the 

wheel bracket and into the clamping plate. 
Using a wrench, tighten the mounting screws 
to secure the wheel bracket to the base. 
Carefully tip the scale to its upright position.  

 

 
 
 
 
 
 
 
 
 
 

To transport the scale, tip the pillar backwards until the scale is resting on the wheel bracket. 
Carefully move the scale to its new location. 

Base 

Mount wheel bracket here 

Close-up view of wheel mounting area 

Clamping plate 

Wheel 

bracket 

Mounting screws 

Mounting screws 

Summary of Contents for 402LB

Page 1: ...P N UM402LB REV20190501 1 402LB Physician Beam Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 402LB Pelstar LLC 2019 0459 ...

Page 2: ...ifications for This Scale 4 Certifications Disposal Service 4 Assembly Instructions 5 Operating Instructions 7 Height Measurement Instructions 9 Transportation 11 Troubleshooting 12 Maintenance and Cleaning 14 Warranty 15 Note This scale has been factory calibrated and does not require calibration prior to use Please register your scale for warranty coverage at www homscales com For User Instructi...

Page 3: ...xercise caution when assembling the scale pillar Assemble and operate the scale per the enclosed user instructions For accurate weighing this scale must be placed on a flat stable surface For accurate weighing verify before each use the proper operation according to the procedure described in this manual Do not use in the presence of flammable or explosive materials If the scale becomes damaged it...

Page 4: ...28 lb 12 7 kg Platform Standing Area Width 10 1 2 26 7cm Depth 14 35 6 cm Height 3 1 4 8 3 cm Certification Descriptions Scale Disposal This Health o meter Professional scale must be disposed of properly as steel waste Follow the national regional or local regulations which apply to you for disposal of steel waste Customer Service Information The 402LB physician beam scale is shipped disassembled ...

Page 5: ... after receipt of the package The following information details what you will find inside the main carton as you unpack the parts for assembly To prevent scratching any components and avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Carefully remove each assembly from the carton and unwrap the packing ma...

Page 6: ...the top of the platform base Insert the studs into the platform base Note If a chair is not available it is recommended that one person holds the platform base while another person handles the pillar 3 Access the underside of the scale base Obtain the four pillar nuts in the carton and place onto the studs on the underside of the scale base Using the provided wrench tighten the pillar nuts to secu...

Page 7: ...poise weights to indicate the patient s approximate weight 2 Have patient step onto the scale Ensure they are positioned directly in the center of the platform Continue to adjust the upper and lower poise weights until the pointer is centered Note It is not necessary for you to wait for the pointer to stop moving in order to read the correct weight As long as the pointer is moving equally within t...

Page 8: ... UM402LB REV20190501 8 OPERATING INSTRUCTIONS CONTINUED Indicator Value reads 27 25 lb 27 lb 4 oz Value reads 100 lb Combined value 100 lb 27 25 lb 129 75 lb Or Combined value 100 lb 27 lb 4 oz 127 lb 4 oz ...

Page 9: ...e headpiece up or down to the estimated height of the patient Then raise the headpiece into its horizontal position To lower height rod below 46 push the tab at the top of the height rod and gently lower the headpiece 2 Ensure that the patient s posture is upright and that the patient s head is straight and level Slide the headpiece down until it rests on the patient s head 3 Determine the patient...

Page 10: ...MEASUREMENT INSTRUCTIONS CONTINUED Height Rod Below 46 3 8 bove 46 3 8 Height Rod Extended A Cm read 176 cm 0 2 cm 176 2 cm Cm read 105 cm 0 2 cm 105 2 cm Inches read 41 3 8 41 3 8 Inches read 69 3 8 69 3 8 Headpiece Indicator Indicator ...

Page 11: ...floor Position the clamping plate inside the base by aligning the holes in the clamping plate and base 3 Position the wheel bracket onto the base by aligning the holes in the wheel bracket with the holes in the clamping plate 4 Insert the mounting screws through the wheel bracket and into the clamping plate Using a wrench tighten the mounting screws to secure the wheel bracket to the base Carefull...

Page 12: ... trig loop in its range of travel Realignment of the beam may be necessary if it is touching Ensure that the upper and lower poise weights are firmly against the shoulder of the beam Adjust the balance ball screw by turning the screw at the left end of the bar Platform rocks excessively or touches base at any corner Base out of alignment Follow the base alignment instructions on the next page Beam...

Page 13: ...erside of the base obtain a sturdy chair and place the top of the pillar with the height rod side facing down on the seat of the chair as shown 3 On the underside of the scale base remove the two black caps from the heel end 4 With the caps removed you can access the alignment screws that allows the heel end of the base to move up or down Use a 7 64 or 1 8 allen wrench to adjust the screws As you ...

Page 14: ...parts are permanently lubricated DO NOT USE OIL OR LUBRICANTS Cleaning and Disinfecting Proper care and cleaning is essential to ensure a long life of accurate and effective weighing 1 Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild soap and water solution 70 isopropyl alcohol solution with 1 5 hydrogen peroxide concentration Afte...

Page 15: ...ding tampering damage in transit or unauthorized repair or alternations Further the warranty does not cover natural disasters such as fire flood hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make sure you keep your...

Reviews: