19
After entering the parameter setting menu, press the "Disp. /
▲
" or "Flush /
▼
" keypad
to select "Alarm setting" White on Black means it has been selected. Press the "Stop /
"
keypad to confirm the settings menu to enter the alarm category including various settings, as
shown below:
PF life
MF life
UV life
Feed
RO
DS
PF
life
400h
500h
600h
:
Alarm time setting for pretreatment failure
In this menu level the maintenance interval for the pre-treatment cartridge ,it must be
reset after exchanged ,the manufacture setting is 400 hour, the user can press
“
Disp./
▲,
Flush/
▼”
keypad to select one of the 400,500,600 hour according to the tap water quality.
Press the “
Stop
/
” keypad to confirms the setting and switches to the next menu level.
Select alarm time UV lamp failure
LCD display is as follows:
PF life
UV life
MF life
DS
UV
4000h
RO
Feed
life :
6000h 8000h
In this menu level the maintenance interval for the UV oxidator of main system must be
reset after exchanging the UV lamp. the manufacture setting is 4000hour for UV lamp life
time, the user can press “Disp./
▲ ,
Flush/
▼ ”
keypad to select one of the
4000,5000,6000,hour according to the user desire. Press the “Stop/
” keypad to confirms the
setting and switches to the next menu level.
Micro-filter failure time alarm options
LCD display is as follows: