HDE DVI-100 Operating Instructions Manual Download Page 3

Kilo voltímetros medidores de alto voltaje alta tensión digitales DVI-100/K02 HD-ELECTRIC manual ingles www.viaindustrial.com

 

 

 

 

SAFETY

 

 

Only

 

trained,

 

professional

 

operating

 

personnel

 

should

 

use

 

the

 

DVI.

 

The

 

voltages

 

this

 

instrument

 

operates

 

at

 

are

 

dangerous

 

and

 

lethal.

 

Severe

 

injury

 

or

 

death

 

can

 

occur

 

if

 

improperly

 

used.

 

 

Risk

 

of

 

electrocution

 

is

 

inherent

 

in

 

or

 

around

 

high

 

voltage.

 

 

Always

 

use

 

proper

 

high

 

voltage

 

procedures,

 

including

 

personal

 

protective

 

equipment,

 

when

 

working

 

near

 

or

 

around

 

high

 

voltage

 

equipment

 

or

 

conductors.

 

 

Do

 

not

 

exceed

 

the

 

DVI

 

maximum

 

voltage

 

rating.

 

 

Assemble

 

the

 

DVI

 

with

 

the

 

proper

 

contact

 

probe

 

for

 

your

 

application.

 

 

The

 

DVI

 

must

 

be

 

used

 

with

 

a

 

hotstick

 

of

 

the

 

appropriate

 

length

 

for

 

the

 

voltage

 

being

 

measured

 

per

 

your

 

company

 

and

 

OSHA

 

published

 

requirements.

 

 

Do

 

not

 

touch

 

the

 

DVI

 

during

 

measurements.

 

The

 

DVI

 

housing

 

should

 

be

 

considered

 

to

 

be

 

at

 

the

 

same

 

voltage

 

as

 

the

 

conductor

 

under

 

test.

 

 

Prior

 

to

 

using,

 

inspect

 

the

 

instrument

 

for

 

any

 

physical

 

damage,

 

cleanliness

 

and

 

check

 

for

 

proper

 

working

 

order

 

by

 

pressing

 

and

 

holding

 

the

 

ON

 

button.

 

Do

 

not

 

proceed

 

if

 

the

 

display

 

does

 

not

 

indicate

 

all

 

8’s

 

or

 

if

 

the

 

beeper

 

does

 

not

 

sound.

 

 

Never

 

allow

 

another

 

high

 

voltage

 

or

 

grounded

 

conductor

 

to

 

contact

 

the

 

instrument

 

during

 

use.

 

Keep

 

the

 

DVI

 

housing

 

free

 

and

 

clear

 

of

 

all

 

structures

 

at

 

all

 

times.

 

Bridging

 

the

 

DVI

 

probe

 

or

 

housing

 

from

 

line-to-ground

 

or

 

line-to-line

 

may

 

cause

 

a

 

fault

 

and

 

arc.

 

 

The

 

DVI

 

does

 

not

 

indicate

 

voltages

 

below

 

100

 

volts.

 

Always

 

use

 

proper

 

grounding

 

procedures.

 

A

 

zero

 

 

These

 

important

 

labels

 

are

 

affixed

 

to

 

the

 

products.

 

Read

 

and

 

understand

 

before

 

proceeding.

 

DVI-100

 

 
 
 
 
 
 
 

DVI-100T

 

 
 
 
 
 
 
 

DVI-500

 

 
 
 
 
 
 
 

DVI-500T

 

voltage

 

reading

 

on

 

the

 

DVI

 

does

 

not

 

mean

 

the

 

line

 

is

 

dead

 

or

 

grounded.

 

 

Grounded

 

equipment

 

or

 

lines

 

can

 

appear

 

to

 

be

 

live

 

when

 

in

 

close

 

proximity

 

to

 

energized

 

conductors

 

if

 

not

 

fully

 

or

 

properly

 

grounded.

 

 

Always

 

test

 

the

 

DVI

 

on

 

a

 

known

 

voltage

 

source

 

before

 

and

 

after

 

each

 

use,

 

or

 

use

 

the

 

PT-DVI

 

Proof

 

Tester.

 

 

The

 

DVI

 

is

 

a

 

voltage

 

indicator,

 

not

 

a

 

voltmeter.

 

Do

 

not

 

attempt

 

to

 

make

 

accurate

 

voltage

 

measurements

 

for

 

phasing

 

or

 

other

 

applications

 

with

 

the

 

DVI.

 

 

The

 

DVI

 

is

 

a

 

direct

 

contact

 

device.

 

The

 

metal

 

portion

 

of

 

the

 

DVI

 

probe

 

must

 

contact

 

the

 

metal

 

conductor

 

to

 

be

 

tested.

 

Voltage

 

indication

 

on

 

insulated

 

conductors

 

such

 

as

 

tree

 

wire

 

or

 

spacer

 

wire

 

will

 

be

 

much

 

lower

 

than

 

actual

 

and

 

may

 

not

 

indicate

 

any

 

voltage

 

at

 

all.

 

 

DVI

 

voltage

 

readings

 

are

 

sensitive

 

to

 

geometry.

 

Read

 

and

 

understand

 

the

 

Accuracy

 

section

 

of

 

this

 

manual

 

before

 

using

 

the

 

DVI.

 

 

The

 

DVI

 

indicates

 

AC

 

voltage

 

only.

 

It

 

does

 

not

 

detect

 

DC

 

voltage

 

such

 

as

 

charged

 

capacitors

 

or

 

cable.

 

3

 

Summary of Contents for DVI-100

Page 1: ...metros medidores de alto voltaje alta tensión digitales DVI 100 K02 HD ELECTRIC manual ingles www viaindustrial com DIGITAL VOLTAGE INDICATORS a n d ACC E S SOR I E S Operating Instruction Manual DVI 100 DVI 500 ...

Page 2: ... 1kV 100 volts to 9 9kVand from 10kVto 99kVfor the DVI 100and to 500kVfor the DVI 500 Anaudible alarm beeps whenever the indicated voltage is0 1kVor higher There are two versions available of each model The DVI 100and DVI 500indicate voltage bydirectly contacting conductors on overhead and underground systems The DVI 100Tand DVI 500Twith TestPoint mode also indicate voltage on elbow test points An...

Page 3: ...igh voltage or grounded conductor to contact the instrument duringuse Keep the DVI housing freeand clearofallstructures at alltimes Bridgingthe DVI probe or housing fromline to ground or line to linemaycause a faultand arc TheDVI does not indicate voltages below 100volts Alwaysuse proper grounding procedures Azero Theseimportantlabelsare affixedtotheproducts Readandunderstand beforeproceeding DVI ...

Page 4: ...before and after each use such as the PT DVIProof Tester Apply the DVIprobe to the conductor to be tested making sure to make metal to metal contact All readings are in kV Voltages below 0 1kV 100V may indicate zero Azero voltage reading does not mean the line is dead or grounded Voltage in excess of 99kV line to ground for the DVI 100 and in excess of 500kV for the DVI 500 will cause the display ...

Page 5: ...higher voltage overbuild accessory hardware such as fuses or disconnects or for measurements made on points such as terminations or endpoints best accuracyiswith the DVI as far away as possible from other liveconductors or grounds awayfrom points or corners and facing the ground Fig 3 4 Expected overall accuracyis 25 For measurements made on grounded conductors keep the Fig 1 Fig 2 Fig 3 DVI as fa...

Page 6: ... where grounded orother liveconductors maybe in closeproximityto the conductor being tested Toassemble the IEP DVI 5probe on the DVI 1 Inspect the IEP DVI 5probe foranymechanical defects and make certain itisclean and dry 2 Screwthe IEP DVI 5into the DVI Makecertain itissnug byhand tightening but DONOTOVERTIGHTEN 3 Testthe fullyassembled DVI on a known voltage source such as the PT DVIProofTester ...

Page 7: ...Displayhousing 4 25 Hx7 Wx4 5 D 11 cm x18 cm x11 cm Overhead probes HP DVI 2 8 long 20 cm for conductors to 1 diameter HP DVI 6 12 long 30cm forconductors to 6 diameter WEIGHT 1 37 lb 0 62 kg with overhead probe LED HEIGHT DVI 100 2 25 6cm DVI 500 1 45 3 7cm ACCURACY Laboratory calibrated to 3 Typicaloverhead line in compact overhead three phase 10 Other applications up to 25 BATTERYLIFE About 100...

Page 8: ...using parts sourced froma varietyofmanufacturers Due to the rapidlychangingtechnologyenvironment partsmaybecomeobsolete unavailableovertime endoflife IntheeventthataProductcannotberepaired orreplacedduetounavailabilityofparts HDElectricCompanywillusecommerciallyreasonableeffortstoobtainsubstitutepartsorconductworkaround design but cannot guarantee itsabilityto do so Itemsnotfounddefectivewillberet...

Reviews: