The linearization software supports up to 6 correction points, though in practice
1 or 2 are almost always enough. The linearization points should be placed at
the points on the weight curve where the greatest variations occur.
Linearization points must be lower than the set maximum load. The lines
between adjacent linearization points are calculated as straight, meaning the
variations are interpolated in linear mode.
Menu
Key
Function/Secondary selection
Select group 1-9
Press 3 x for group
3 Linearization
→ Linearization points already present
Fixpoint 1:
999999
Display existing linearization points (1 to max. 6)
Delete existing linearization point
Calibrate Fixpt?
N
Lin.Signal
0.50000
Linearization signal is displayed
→ Clr key pressed in
Fixpoint n
step
Delete Fixpoint n? N
Y
Delete linearization point; continue to
New
Fixpoint?
step
N
Do not delete linearization point; continue
to next linearization point
Information
After deleting a linearization point, the remaining points are renumbered in
ascending order.