Structure and mode of operation
T10F
A0608-14.0
HBM: public
15
5
Structure and mode of operation
Torque flanges consist of two separate parts: the rotor
and the stator. The rotor comprises the measuring body
and the signal transmission elements.
Strain gauges (SGs) are mounted on the measuring
body. The rotor electronics for transmitting the bridge
excitation voltage and the measurement signal are
located centrally in the flange. The transmitter coils for
contactless transmission of excitation voltage and
measurement signal are located on the measuring body's
outer circumference. The signals are sent and received
by a separable antenna ring. The antenna ring is
mounted on a housing that includes the electronic
system for voltage adaptation and signal conditioning.
Connectors for the torque signal, the voltage supply and
the speed signal (option) are located on the stator. The
antenna ring should be mounted more or less
concentrically around the rotor (
In the case of the speed measuring system option, the
speed sensor is mounted on the stator, the customer
attaches the associated slotted disc on the rotor. The
speed measurement works optically with the infrared
transmitted light barrier principle.