Electrical connection
12
A03313_02_Y00_00
HBM: public
PW29
5
Electrical connection
The following can be connected for measurement signal conditioning:
S
Carrier‐frequency amplifier
S
DC amplifier
designed for strain gage measurement systems.
In the standard version, the transducers come supplied with a 3m cable with a
6‐pin Pancon plug in a six‐wire configuration.
5.1
Connection with a six‐wire configuration
4
Plug‐in contact 1 (white) =
measurement signal (+)
Plug‐in contact 3 (black) =
excitation voltage (-)
Plug‐in contact 6 (gray) =
sense lead (-)
Plug‐in contact 4 (blue) =
excitation voltage (+)
Plug‐in contact 5 (green) =
sense lead (+)
Plug‐in contact 2 (red) =
measurement signal (-)
Schematic diagram of a Pancon plug (CE100F26‐6), 6‐pin
1
2
3
5
6
blue marking
Shield (yellow) = cable shield
connected to load cell body
Fig. 5.1
Connection with a 6‐wire cable (choice of 3m or 6m lengths)
With this cable assignment, the output voltage at the amplifier is positive when
the transducer is loaded (
).
5.2
Connection with a four‐wire configuration
When transducers with a six‐wire configuration are connected to amplifiers with
a four‐wire configuration, the sense leads of the transducer must be connected
to the corresponding excitation voltage leads: (+) with (+) marking and (-) with
Summary of Contents for PW29 Series
Page 1: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Français PW29 ...
Page 3: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Français PW29 ...
Page 19: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Français PW29 ...
Page 37: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Français PW29 ...
Page 55: ...Dimensions PW29 A03313_02_Y00_00 HBM public 19 ...