PW15AHi
12
A2526-4.0 en/de/fr
HBM
Users can set their own characteristic curve (commands
LDW, LWT, NOV
)
without changing the working standard calibration. Gross/net selection is also
available (
TAS
command). Command
ZSE
can be used to activate an
automatic zero on start‐up function. There is also an automatic zero tracking
function (
ZTR
) available.
Command
LIC
is available for linearization of the scale curve (with a third order
polynomial). The polynomial parameters can be set using the HBM PC
program, AED_Panel32.
The current measured value is read out using the
MSV?
command. The format
of the measured value (ASCII or binary) is set with the
COF
command. You
can also use the
COF
command to select automatic measurement output.
Different types of digital filter are implemented in the PW15AHi, and these are
selected with the
FMD
command. With
FMD0
, filters are also available below
the 1 Hz limit frequency. In filter mode
FMD1
, fast‐settling filters with high
attenuation in the stopband are activated. For more detailed information, see
the Online Help.
5.3
Trigger functions
The PW15AHi includes
four trigger functions
to support measurements in
packaging machinery and checkweighers:
Internal pre‐triggering via an adjustable level
External triggering via the digital control input (Pre‐Trigger)
Internal post‐triggering via an adjustable level
External triggering via the digital control input (Post‐Trigger)
Gross or net measured values can be conditioned as measured values (
TAS
).
The input mode should be set to
IMD
1.
The trigger function is described in the Online Help, Trigger Function Overview.