Parts in contact with the medium are made of 1.4301 and 1.4542 stainless
steel. Any fluid or gas (vapour) that does not attack these steels is a suitable
measurement medium. The pressure transducer is attached by means of its
threaded connector for the pressure connector and can be installed in any
position. In particular cases there are specific instructions in Chap. 4
”Assembly” which must be followed!

3

Construction and operating mode

The internal measurement tube bears the strain gauges that are connected to
a Wheatstone bridge. The SG application is located in a hermetically sealed
reference chamber. This protects it from ambient effects and the pressure
measurement is not affected by the ambient pressure (absolute pressure
transducer).

The measuring membrane is strained by the pressure applied. As a result,
with thin-film strain gauges the ohmic resistance changes proportionally to the
pressure and the Wheatstone bridge is detuned.

4

Installation

Caution: Pulling on the housing or the cable entry will result in
mechanical damage to the transducer.

Pressure transducers can be screwed into place in any location. If a
transducer is used for measuring dynamic pressure trends in fluids, it must
be installed with the pressure connector upwards so that an air-bulge cannot
form in the measurement tube.

When tightening, the wrench (size 27) must only be applied
to the clamping area of the mounting flange, and not to the
housing or cable entry
The permitted tightening torque is
30Nm.

WARNING

SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:[email protected]

SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:[email protected]

Summary of Contents for P6A

Page 1: ...g instructions Electrical measurement of mechanical quantities SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com SUNSTAR http www sensor ic com TEL 0755 83376489...

Page 2: ...SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com SUNSTAR http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 E MAIL szss20 163 com...

Page 3: ...5 1 Cable extension 12 6 Differential pressure measurement circuit 13 7 Measurement 14 7 1 Measuring dynamic pressures 14 8 Technical Data in accordance with DIN 16 086 15 9 Dimensions in mm 16 10 Op...

Page 4: ...technically correct storage installation and assembly coupled with careful operation General dangers in non observance of the safety instructions The pressure transducer corresponds to the state of t...

Page 5: ...is Manual with the following symbols Symbol DANGER Meaning Maximum danger level Signifies a directly dangerous situation which if the safety requirements are not observed will result in death or serio...

Page 6: ...equirements of the relevant EC guidelines see Certificate of Conformance at the end of this User Manual Safety conscious working Fault messages must only be acknowledged if the cause of the fault has...

Page 7: ...data relating to the safety requirements and safety rules and regulations below It is also essential to comply with the statutory and safety regulations relevant to the particular application This als...

Page 8: ...and corrosion resistant steel and can be used from 30 to 100 C 2 Corresponds to PT06E 10 6S supplied by Bendix or UPT06J 10 6S supplied by Canon 2 Application Absolute pressure transducers are suitab...

Page 9: ...ansducer The measuring membrane is strained by the pressure applied As a result with thin film strain gauges the ohmic resistance changes proportionally to the pressure and the Wheatstone bridge is de...

Page 10: ...against the transducer and the bearing surface Caution Before the P6A pressure transducer is installed or removed care must be taken to ensure that there is no pressure in the circuit AF 27 USIT Ring...

Page 11: ...inated connector cable and option with HS6P connector Pin assignment is shown in Table 5 1 The cable shielding is connected to the transducer housing see Greenline shielding design Pin assignment Stan...

Page 12: ...e and spacer until the soldering tags and cable shielding are accessible When dismantling the cable socket please do not pull on the white jack assembly 5 1 Cable extension Use shielded low capacitanc...

Page 13: ...ge contacts are connected in parallel and the measurement voltage contacts are connected in crossover The resulting resistance is then 175W 1 2 3 4 2 bk 4 rd 3 bu 1 wh Transducer 1 Transducer 2 1 2 3...

Page 14: ...ally Dead volume and any fluid or gas volume present on the user side can also lead to a false measurement result 7 1 Measuring dynamic pressures Calibration by reference to static pressures also appl...

Page 15: ...ivity mV V 2 Sensitivity tolerance variance of sensitivity from rated sensitivity 0 2 Influence of temperature on sensitivity in the rated range of the excitation voltage per 10K with reference to act...

Page 16: ...ring signal WH measuring signal ground GE not used GN Assignment of the connector optional K L Size R D M N D N M BK excitation voltage BL excitation voltage Size P6A D K L M N R Size with cable conne...

Page 17: ...D K L M N O P R S P3M 500 M20 M12x1 5 25 50 M20x1 5 4 20 2 5 5 3 P3M 500 R1 2 M12x1 5 20 50 G1 2 4 20 2 5 5 3 SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com S...

Page 18: ...ypical order for 10 bar with D15 connector 1 Please state length of cable required 0 1 0 B D Typical order P6A 2 Corresponds to PT06E 10 6S supplied by Bendix UPT06J 10 6S supplied by Cannon Order num...

Page 19: ...opy of Declaration of Conformity SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com SUNSTAR http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 E MAIL szss...

Page 20: ...20 P6A SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com SUNSTAR http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 E MAIL szss20 163 com...

Page 21: ...21 P6A SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com SUNSTAR http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 E MAIL szss20 163 com...

Page 22: ...22 P6A SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com SUNSTAR http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 E MAIL szss20 163 com...

Page 23: ...23 P6A SUNSTAR http www sensor ic com TEL 0755 83376549 FAX 0755 83376182 E MAIL szss20 163 com SUNSTAR http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 E MAIL szss20 163 com...

Page 24: ...difications reserved All details describe our products in general form only They are not to be understood as express warranty and do not constitute any liability whatsoever im d6 97 0pod e SUNSTAR htt...

Reviews: