15
P6A
8
Technical Data (in accordance with DIN 16 086)
Type
P6A
Accuracy class
0.2
Measuring range, 0 bar -
bar
10
20
50
100
200
500
Natural frequency of the membrane,
approx.
kHz
13
15
26
38
65
>100
Mechanical values in accordance with
VDI/VDE 2600,
by reference to final scale value
Operating range at +23
C
%
0...200
0...150
Overload cutoff at +23
C
%
250
200
Test pressure
%
250
200
Destructive range
%
>250
>200
Materials of the
internal interface (threaded
connector/measuring chamber)
1.4301/1.4542
1.4542
external interface (threaded
connector/housing)
1.4301, 1.4541, nickel-plated
brass, polyamide,
silicone rubber (cable)
1.4542/
1.4541
Rated range of the excitation voltage
V
0.5...12
Input resistance at +23
C
W
350
5
Output resistance at +23
C
W
350
5
Rated sensitivity
mV/V
2
Sensitivity tolerance
(variance of sensitivity from rated
sensitivity)
%
<
0,2
Influence of temperature on sensitivity in
the rated range
of the excitation voltage per 10K, with
reference to actual value
in the rated temperature range
in the operating temperature range
%
%
<
0,1
<
0,2
Influence of temperature on the zero
signal in the rated range of the excitation
voltage per 10K, with reference to rated
sensitivity
in the rated temperature range
in the operating temperature range
%
%
<
0,15
<
0,2
Linearity variation including hysteresis
%
<
0,2
Repeatability in accordance
with DIN 1319
%
<
0,05
Rated temperature range
C
-10...+80
Service temperature range
C
--40 ... +100 (... +120
C up to 24 hours)
Storage temperature range
C
--40 ... +100 (... +120
C up to 24 hours)
Weight without cable, approx.
g
200
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:[email protected]
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:[email protected]