12.2.4
GN611, Isolated 1 kV 200kS/s input card
l
6 analog channels
l
Isolated, balanced differential inputs
l
±
20 mV to
±
1000 V input range
l
600 V RMS CAT II reinforced isolation, tested up to 6.4 kV
l
Analog/digital anti
-
alias filters
l
200 kS/s sample rate
l
18 bit resolution
l
200 MB memory
l
Two 4 mm banana plugs for each channel
l
Real
-
time cyclic calculators
l
Triggering on real
-
time results
l
Digital Event/Timer/Counter support
The isolated balanced differential inputs offers voltage ranges
from
±
20 mV to
±
1000 V.
Tested up to 6.4 kV, the reinforced isolation allows for safe
measurements up to 600 V RMS CAT II (IEC61010
-
1:2010 safety
standard).
Optimum anti
-
alias protection is achieved by the 7
-
pole analog
anti
-
alias filter combined with a fixed 2 MS/s sampling Analog
-
to
-
Digital converter. At lower sample rates the digital anti alias filters
allow for a large range of high order filter characteristics with
precise phase match and ultra low noise output.
Besides the instantaneous values, the real
-
time cycle detection allows for cycle
based real
-
time calculations like TrueRMS on all analog channels as well as
torque, angle and speed on all timer counter channels simultaneously.
Every real
-
time calculated result can be used to trigger the recording.
If supported by the selected mainframe, the GEN DAQ series input card offers
16 digital input events, two digital output events and two Timer/Counter
channels.
The two Timer/Counters together with the GEN series mainframe Digital Event/
Timer Counter connector and the G070A torque/RPM adapter allow for direct
interfacing to HBM torque transducers or other torque and speed sensors.
For specification and ordering information, please refer to "B3716-4.2 en (GEN
series GN611)" on page 549.
1
2
!
2MΩ 1kVrms
CAT II 600V
3
!
2MΩ 1kVrms
CAT II 600V
4
!
2MΩ 1kVrms
CAT II 600V
5
!
2MΩ 1kVrms
CAT II 600V
6
!
2MΩ 1kVrms
CAT II 600V
!
2MΩ 1kVrms
CAT II 600V
GEN3i
214
I3763-3.1 en HBM: public