Command set of the HBM interpreter
Amplifier settings
46
A03474_04_E00_00
HBM: public
DMP41
Effect:
The input value is saved in the amplifier zero memory.
Response:
Acknowledge-
ment
Significance
0
Command executed
?
Error
Example 1:
Start zero setting
CDW(x)
0
(y)
Example 2:
CDW10000(x)
0(y
)
Information
If the value read out with CDW?1 (absolute value) is sent for p1, the
momentarily present gross measurement signal is set to zero.
CDW7680000(x)
0(y)
Zero value set to 7680000 ADU
CDW2.5,11(x)
0(y)
Zero value set to 2.5 mV/V
CDW780.75,12(x)
0(y)
Zero value set to 780.75 in scaled unit
Important
It is still possible to set a basic misalignment to zero by calculating the zero
value to be input using the following calculation:
Zero value (ADU units) =
7680000 x basic misalignment (mV/V)
Measuring range full scale (mV/V)
CDW: If correspondingly programmed, the input value is inverted