DIS2116
42
HBM
A2540-1.1 en/de
Maximum cable lengths
1)
between DIS2116 and C16i load cell(s)
If the supply voltage of the DIS2116 is less than 17 V, the digital load cells can
be directly supplied from the scale electronics. Please note the below table for
dimensioning. With a nominal (rated) supply voltage of 24 V, the load cells re-
quire a separate voltage supply.
Nominal (rated) voltage
12 V
DC
Nominal (rated) voltage
15 V
DC
Core cross section (mm
2
)
0.14
0.25
0.5
0.14
0.25
0.5
Number
of load
cells
Total Power
consump-
tion (mA)
Cable length (m)
1
50
393
500
500
500
500
500
2
100
197
350
500
315
500
500
3
150
131
234
468
210
375
500
4
200
98
176
351
157
281
562
6
300
66
117
234
105
187
375
8
400
49
88
176
79
140
281
10
500
39
70
140
63
112
225
12
600
33
59
117
52
94
187
16
800
25
44
88
39
70
140
1)
Cable length up to max. 500m.