Electrical connection
24
A1460-4.0
HBM: public
C16A
7.5
EMC protection
Background information
Electrical and magnetic fields often induce interference
voltages in the measuring circuit. For reliable measure
ments, signal differences of less than 1μV must be trans
mitted without interference from the transducer to the
evaluation electronics.
Planning of the shielding concept
Due to the numerous application options and the different
local constraints, we can only provide you with informa
tion for correct connection. The shielding concept suit
able for your application must be planned locally by an
appropriate specialist.
The load cells with shielded, round cables are EMC-
tested in accordance with EC directives and identified by
CE certification.
Points to be observed
S
Use shielded, low-capacitance measurement cables
only (HBM cables fulfill both conditions).
S
Do not route the measurement cables parallel to
power lines and control circuits. If this is not possible,
protect the measurement cable with e.g. steel con
duit.
S
Avoid stray fields from transformers, motors and con
tact switches.
S
Connect the shield of the connection cable
extens
ively
on the shielding electronics housing. When using
several load cells, connect the shield extensively to
the junction box (combination of transducer signals,
e.g. type VKK2 from HBM). From there, connect the
Summary of Contents for C16A C3
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais C16A...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais C16A...
Page 41: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais C16A...
Page 82: ...Abmessungen 42 A1460 4 0 HBM public C16A...
Page 83: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais C16A...