Specifications
BN100A
A0643-6.1
HBM: public
15
5
Specifications
Type
BN100A
Accuracy class
0.0005
Supply voltage
V
AC voltage 230, 50 Hz
AC voltage 115, 60 Hz
Degree of pollution,
max. altitude for operation
m
2
2000
Overvoltage category
II
Power consumption
VA
20
Calibration signal
mV/V
-100...+100
Steps
mV/V
0.1
Calibration error
1)
related to 2 mV/V
%
t
0.0005
Step error
, related to the specific step
value
%
0.0003
Calibration signal deviation on chan
ging polarity,
related to 2 mV/V
%
0.0004
Output noise voltage in permissible
frequency range
nV/
Ǩ
Hz
t
20
Nominal temperature range
o
C
+15...+30
Permissible ambient temperature
range
o
C
0...+50
Storage temperature range
o
C
-20...+60
Nominal value of the bridge excita
tion voltage
(rms value)
V
10
Permissible bridge excitation
voltage
2)
(rms value)
V
1...15
1)
Related to socket Bu1 of the BN100A, with the measuring instrument connected in a 6-wire
circuit.
2)
Traceable to PTB; typically at 2.5 V, 5 V and 10 V