11
HLC / BLC / ELC
HBM
A11303.1en/de/fr
Wi
r
ing code
6
wi
r
e ci
r
cuit
(white)
(black)
(grey)
Excitation (+)
(green)
(red)
(yellow)
Excitation (−)
Sense (−)
Signal (+)
(blue)
Sense (+)
Signal (−)
Shield, connected with housing
W
hen connecting 6wire load cells to amplifiers in 4wire technique you must
connect the wires
blue to g
r
een
and
black to g
r
ey
are to be connected up.
The following deviations occur on
unsho
r
tened
cables (3m):
S
Sensitivity (C
n
) −0.2%
S
TK
C
−0.01% / 10 K
Wi
r
ing code
4
wi
r
e ci
r
cuit
(only applies to versions with 4wire connection cable)
(white)
(black)
Excitation (+)
(green)
(red)
(yellow)
Excitation (−)
Signal (+)
Signal (−)
Shield, connected with housing
C
able extension
For extending cables, only use shielded low capacity measurement cables
(measurement cables provided by
HBM
1)
meet these requirements). Ensure
that there is a perfect connection with the lowest possible transition
resistance. If you use the 6wire circuit, you may neglect any interference by a
change in the resistance of the extension cables. However, if you extend the
cable using the 4wire technique, the characteristic value deviation can be
removed by adjustment. Temperature influences will not be compensated.
1)
e.g.
HBM
extension cable, 6 wires:
•
KAB8/002/2/2 (sold by the meter, order no.
43301.0071
= grey or
43301.0082
= blue)
•
CABA1 (cable reel, order no.
1CABA1/20
= 20m or
1CABA1/100
= 100
m in length), grey
•
CABE2 (cable reel, order no.
1CABE2/20
= 20m or
1CABE2/100
=
100m in length), blue