AED9101D
I3563-1.0 en
17
AED9101D
HBM color code:
BU = blue
GN = green
RD = red
WH = white
GY = gray
BK = black
GN (3')
RD (4)
WH (1)
GY (2')
BK (2)
BU (3)
TE1
TE4
1
2
16
1
15
7
Fig. 6.3
Connecting transducers with a fourwire
configuration without an extension cable
Plug connection
Cable to the transducer
electronics
BU
RD
WH
BK
BU
GN
RD
WH
GY
BK
Cable shield
Transducer
Fig. 6.4
Connecting transducers with a fourwire
configuration via a 6wire extension cable
When transducers are in a four-wire configuration, the
connections for the sense leads must be connected to