Assembly Instructions M8155-EN 

 

M8155-EN   2018-03 

 

3/5

 

 

  

     

 The initialisation window (Fig. 2) is displayed at the individual 

device or operation module as soon as a new DFM-53 was 
detected. 

 

 

 

 

 

Fig. 2: Assignment DFM-53 

7.

 

Proceed as follows to set the affiliation of a flow rate meter 
board DFM-53 to a device:  

  Assign flow rate meter board ID (left column) to a unit by 

entering the device address (right column). 

  Acknowledge assignment with the 

 key. 

  Exit initialisation window using the 

 key. 

 

 

 

NOTICE! 

Failure to assign the DFM-53 will render the unit 
unusable. 
The affiliation can be subsequently set via 

Service \ 

Configuration \ Assignment DFM-53

. The 

Service \ 

Konfiguration

 menu is accessed via the 1?2? code. 

 

  

 

 

 

8.

 

Main switch off and switch on again after 30 s

 

 

9.

 

Switch the unit on with the 

 key. 

   

 

 

Fig. 3: Calibrating the flow rate DFM-53 

10.

  Flow rate calibration 

  Operate unit in regular mode at 40 °C for at least 10 minutes. 

  Set 

Pressure relief with unit OFF

 parameter in 

Setting \ 

Miscellaneous

 to "OFF". 

  Switch off unit using 

 button and wait at least 10 seconds. 

  Set 

Flow rate calibration DFM-53

 parameter in 

Service / 

Calibrating / Flow rate

 internally to "ON". 

  The flow rate is calibrated automatically. 

  Set 

Pressure relief for unit OFF

 parameter in 

Setting \ 

Miscellaneous

 to "ON". 

   

 