Assembly Instructions M8125-EN 

 

M8125-EN   2018-03 

 

4/5

 

 

  

 

 

 

 

 

Fig. 5: Calibrate flow rate during modular 
operation 

Operation Modular 

8.

 

Proceed as follows to calibrate the corresponding flow rate 
metre: 

  In normal operation operate the device at 40 °C for at least 10 

minutes. 

 

Only in the case of water devices:

 

If present set the parameter 

Pressure relief with unit OFF

 at 

Setting / Miscellaneous

 to "OFF". 

  Switch the unit off using button 

 and wait at least 10 

seconds. 

  Select the module affected e.g. "A1" with the 

 or 

 

buttons. 

  Set parameter 

Flow rate calibration

 at 

Service / Calibration / 

Flow rate

 to "ON". 

  The flow rate is calibrated automatically. 

 

Only in the case of water devices:

 

If present set the parameter 

Pressure relief at unit

 

OFF

 at 

Setting / Miscellaneous

 to "ON". 

 

Switch the unit on with the 

 key. 

   

 

 

9.

 

Inspection of calibration of Flow rate meter 

  In normal operation operate the unit at 40 °C for at least 5 

minutes following calibration. 

  Close the corresponding shut-off valve between main and 

return lines. 

  Flow rate returns to 0 L/min. 

  If this is not the case repeat the Flow rate meter point. 

 

 

  Punkt Durchflussmessung kalibrieren für jeden ersetzten 

Schallwandler wiederholen. 

 

 

10.

  Switch the unit off by press the 

 key. 

 

11.

  Main switch off. 

  

 