24
2
Operation
Capacitance measurement / F
Set the selector switch to "
/ ". Actuate the "
SEL
" button x 3. Connect the black test
lead to the "
COM
" jack and the red test lead to the "
V/Ω
" jack. Connect the test leads with the
test object and read off the measurement result from the display.
Measurement range
Resolution
Accuracy
60 nF
10 pF
± 4 % + 3 digit
600 nF
100 pF
6 µF
1 nF
60 µF
10 nF
600 µF
100 nF
6 mF
1 µF
60 mF
10 µF
Discharge the capacitors before every measurement!
Automatic switch-off
The instrument switches off after c. 15 minutes (APO). Continuous actuation
of the "
APO
" button switches the Auto Power Off function on or off.
Backlighting
Continuous pressing of the "
HOLD
" button switches on the backlighting.
The illumination switches off automatically after c. 1 minute.
Saving function (HOLD)
At difficult measurement positions, the display cannot perhaps be read correctly.
Pressing the "
HOLD
" button freezes the current measured value which can then
be read in comfort.
Repeated pressing of the "
HOLD
" button returns to normal measuring mode.
True RMS
If measuring non-sinus wave forms, low measurement errors can occur if using the True RMS
function, in comparison to using traditional measurement procedures.
Sinusform and non-sinus form signals can be accurately measured using the True RMS function.
^