101B - PAGE 11
4.0 OPERATION OF VACUUM GAUGE
1
) Plug the 6-foot power cable into any single 115V phase AC line. Line frequency is unimportant and 50, 60 or 400 cycle
power will provide satisfactory operation. A 3-wire grounded power cord is supplied with all gauges, including the European
230V models (prefix "E" added to the model number denotes 230V, i.e. EVT-4). Where applicable, turn the ON-OFF switch to
"ON". Allow several minutes for warm-up.
NOTE: VT-4A, 5A, and 6A Models require 115 VAC; EVT-4A, 5A, and 6A Models require 230 VAC, 50-60 cycle, single
phase supply voltage.
2
) With the gauge tube unplugged, set the mechanical zero to indicate the center dot past "ATM" for all VT-6 models (see Fig.
5.1, page 9).
3
) Plug gauge cable onto gauge tube, and meter should indicate "ATM". If a calibration check is desired, follow the directions
in Section 5.2.1 or 5.2.2, whichever is applicable.
4
) Gauge can now be operated with a vacuum tube that has been installed as detailed in Section 3.5.
Summary of Contents for DV-20
Page 2: ...101B PAGE 2 ...
Page 16: ...101B PAGE 16 ...
Page 18: ...101B PAGE 18 ...
Page 20: ...101B PAGE 20 ...
Page 21: ...101B PAGE 21 ...
Page 22: ...101B PAGE 22 ...
Page 23: ...101B PAGE 23 ...
Page 24: ...101B PAGE 24 ...
Page 25: ...101B PAGE 25 ...
Page 26: ...101B PAGE 26 VT 4B VT 6B Schematic information from Hastings drawing 30429 rev P ...
Page 27: ...101B PAGE 27 VT 4S VT 6S Schematic ...
Page 29: ...101B PAGE 29 VT 4A VT 5A VT 6A Schematic information from Hastings drawing 30472 rev J ...
Page 30: ...101B PAGE 30 VT 4S2 VT 6S2 Schematic information from Hastings drawing 30506 rev F ...
Page 31: ...101B PAGE 31 ...
Page 32: ...101B PAGE 32 ...
Page 33: ...101B PAGE 33 ...
Page 34: ...101B PAGE 34 ...