page 15
3.7.2 Calibration Using A Hastings Reference Tube
Connect a Hastings reference tube to a stabilized vacuum gauge. Wait until the displayed pressure
is stable.
NOTE:
The following table specifies the Hastings reference tube to be used in the calibration of a
gauge based upon the type of gauge tube being used:
Reference tube
Gauge tube
DB-16D
DV-4
DB-18
DV-5
DB-20
DV-6
Adjust the gauge CAL control (figure 3.2) until the meter pointer indicates the pressure specified
on the decal of the reference tube in use.
figure 3.2; VT/CVT calibration points