Cleaning the Cooling Tank
Cleaning Non-Recirculating Water
Cooling Tank
The non-recirculating water cooling tank should
be drained and rinsed out every six months. Do
not clean the tank when the sterilizer is in
operation.
Using the optional overflow tube (supplied)
push the connector fitting into the drain connec-
tion at the bottom of the tank to empty it.
Remove the drain tube when the tank is empty,
and rinse the tank with a solution of household
bleach (3/4 cup) and distilled or deionized water
(1 gal.). Drain solution and rinse with distilled or
deionized water before refilling.
Refill the tank with clean distilled or low quality
deionized water (minimum 0.5 megohm/cm,
maximum 1.5 megohm/cm) to put the system
back in service.
Hot Surface
Burn Hazard:
Do not clean the
tank when the sterilizer is in
operation. Allow the Non-
recirculating Cooling Tank to cool
to room temperature
before
cleaning the tank.