Control Description
Front Panel
A schematic of the face panel of the
CL-100
is shown below. Important
components are identified by letter (A-H, J-N). Refer to this diagram as an
aid in identification of the described components.
Power Switch (Item A)
Supplies power to the
CL-100
.
Meter Selector Switch and Meter (Item B)
THE SELECTOR SWITCH selects the parameter displayed on the associated
METER. Included options are
Set Temp, Control Temp, Monitor Temp
, and
Output Volts
and are described below.
Set Temp
reports the adjustment from either the SET TEMPERATURE con-
trol (Item H) or the EXT TEMP SET BNC (Item Q). Displayed units are °C.
Control Temp
reports the actual temperature of the feedback thermistor sit-
uated within the heater/cooler device being used. The data is informative
only and is in units of ºC.
Monitor Temp
reports the temperature of the sensor thermistor which con-
nects via the MONITOR TEMP IN BNC (Item F).This provides a convenient
means to determine the temperature of any point of interest within your
setup. Displayed units are ºC.
Output Volts
reports the adjustment from either the MANUAL VOLTAGE
control (Item K) or the EXT VOLT SET BNC (Item P). Displayed units are V.
The METER also provides an indication that the main POWER is
on
.