07/19/00
High-Speed Resistance Tester – Model 9250-05
Operators Manual
12
All Set Averaging values from 1 to 60 will output at 60Hz with Line SYNC set ON. Averaging
values higher than 60 will result in update rates lower than 60 Hz with Line SYNC set ON.
Line SYNC OFF:
This command will turn the power line synchronization off. Line SYNC ON is the factory default
for the Model 9250-05.
Limit Settings
High and low limit settings are available with the Model 9250-05. The Set UPPER Limit command
sets the high limit and the Set LOWER Limit command sets the low limit.
Set UPPER Limit:
Enter the UPPER Limit in OHMS, with the appropriate decimal point. When resistance =
1.5000
and the high limit = 1.4900
the “GO” relay will be open.
Example:
Set UPPER Limit to 20.500
.
Command> *E 20.500
Set LOWER Limit:
Enter the LOWER Limit in OHMS, with the appropriate decimal point. When resistance =
1.0999
and the low limit = 1.1000
the “GO” relay will be open.
Example:
Set LOWER Limit to 10.500
.
Command> *M 10.500
Start Readings:
The Start Readings command will initiate the sending of resistance readings for the current range.
The measured data will be in the following format (this reading is for Range One):
.XXXXX
Note that the decimal point will move, depending upon which range is being tested (i.e. Range
Two = X.XXXX, for Range Three = XX.XXX, etc.).
HELP Screen:
or
This displays a screen with all of the command sequences listed with short descriptions and value
ranges. Also displayed is a short summary of many of the settings for the system under Current
Settings.