76
Chapter
5
•
•
•
•
•
•
Span Weight Parameter
The Span Weight value is the weight of the object that is being placed on the scale to
set the “High” calibration point with respect to the “Low” calibration point set using
the Cal Lo Weight parameter.
On the
Web page
, select the calibration menu and left click inside the Span Weight
text field and enter the load cell sensitivity value.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the Hard Cal submenu, then select the Span weight menu item. The
current Span Weight value will be displayed, press the ENTER or the RIGHT button if this
value needs to be modified.
Cal Date Parameter
The Cal Date parameter is used to record the date when the last calibration was completed
on the HI 6500 series instrument. The format is
mm/dd/yyyy
On the
Web page
, select the calibration menu and left click inside the Cal Date text field
and enter the date the calibration was completed.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the Cal Date submenu. The last Cal Date value will be displayed,
press the ENTER or the RIGHT button if this date needs to be modified.
Multi-Point Calibration
For details on Multi-Point Calibration, see page 123.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......