Overview
3
•
•
•
•
•
•
WAVERSAVER
®
When measuring small weight changes, the effects of mechanical vibration and noise from
feeders and other plant environmental conditions can introduce substantial interference.
WAVERSAVER factors out vibration, noise, and other interference-related signals from the
load cell so the weight processor can better decipher the actual weight data.
WAVERSAVER can factor out noise with frequencies as low as 0.25 Hz or as high as 7.5
Hz. Five cut-off frequencies can be selected, with higher frequencies providing a faster
response time. The default factory setting is 1.00 Hz vibration frequency immunity.
WAVERSAVER
®
+
(HI 6500-XP and HI 6510-XP only)
WAV is a superset of WAVERSAVER for the XP Extreme Processor series that
uses two additional parameters,
Variation Threshold
and
Motion Threshold,
to configure
an adaptive filtering algorithm, which significantly improves the stability and accuracy of
the weight reading.
Weighing System Tests
These tests are used to diagnose drifting or unstable weight reading problems. It requires a
Hardy IT Junction Box for full utilization, for example the HI 215IT or the HI 6010IT. The
ability to read the weight seen by the individual load sensors allows you to use this test for
making cornering, leveling and load sharing adjustments to the weighing system.
INTEGRATED TECHNICIAN provides the following problem detection support:
1.
Integrated Technician Weight and Voltage Test
: Reads the weight of each load
sensor in the system to see if the load sensor might be causing the problem.
2.
Integrated Technician Stability Test
: Disconnects the load sensors and engages
an internal (in the junction box) reference signal to see if the Junction box, the cable
between the instrument and the Junction Box, or the instrument is causing the
problem.
3.
Integrated Technician Return to Zero Test
: Measures and stores the mV out of
each Hardy load cell during a C2 calibration when zero is used as the reference.
Later when the test is run it measures and compares the voltage with zero weight out
from each load cell to the stored value and indicates either a pass or fail. This test
insures the integrity of the weighing system.
NOTE
C2, INTEGRATED TECHNICIAN and WAVERSAVER are registered trademarks of
Hardy Process Solutions.
Using the Set-Up Wizard
The Set-up Wizard is designed to guide you through the basic steps of setting up an HI 6500
instrument for the first time or to easily make changes to an instrument after the initial set-
up is complete. The Wizard can be launched both through the Webserver and the front panel
display (see illustrations below).
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......