Chapter 4 - Setup
4-25
Setting The Zero
Tolerance Value
Sets the range of weights so that the Zero Command
works, as an offset of the calibrated Zero.
Setting the Auto
Zero Tolerance
Value
When the Auto Zero Tolerance is entered, and Auto
Zero Tracking is enabled, any weight within the
entered tolerance of zero and not in motion, will cause
the display to automatically read zero.
NOTE:
There is a short time delay (at least 1 second) before
the AutoZero Triggers.
Setting the
Number of
Readings
Averages
The Number of Averages sets the number of weight
readings which will be used to compute the displayed
weight. The average is a sliding average so that a new
average reading is available for display at every read-
ing.
Setting the Span
Weight Value
The Span Weight is a reference point derived from an
actual measured weight. This should not be confused
with the scale capacity. If you have a 100 pound
weight and you place it on the scale, the Span Weight
is 100 pounds.
Setting the
WAVERSAVER
Value
There are 6 selectable levels, 0 means that WAVER-
SAVER has not been selected. Default setting is #3 1
Hertz.
•
0 = NO WAVERSAVER
•
1 = 4.0 Hertz
•
2 = 2.0 Hertz
•
3 = 1.0 Hertz - Default
•
4 = 0.5 Hertz
•
5 = 0.25 Hertz
Summary of Contents for HI 1769-2WS
Page 14: ...HI 1769 WS HI 1769 2WS WEIGH SCALE MODULE 1 6 ...
Page 27: ...Chapter 3 Installation 3 9 FIG 3 7 RAIL FASTENERS IN THE CLOSED POSITION ...
Page 60: ...HI 1769 WS HI 1769 2WS WEIGH SCALE MODULE 4 26 ...
Page 74: ...HI 1769 WS HI 1769 2WS MANUAL 6 8 ...
Page 82: ...HI 1769 WS HI 1769 2WS MANUAL Weighing System Tests 1 4 wrist strap grounding device 3 2 ...