Chapter 6
C2 Calibration Using Ladder Logic
Verify that the parameters have been setup for your weighing process in accordance with
the setup information provided in Chapter 4.
A Ladder Logic Example is located at the Hardy Process Solutions Inc. Web Site. See
http://www.hardysolutions.com
. Click products > PLC Weighing Modules > HI 1756-
FC. Click on the Docs & Programs tab and scroll down to Programs and Firmware.
Select
HI 1756-FC Sample Programs
for the Ladder Logic example
Hard Calibration
Hard Calibration is the traditional method of calibration using test weights. Hardy
recommends that the test weights total 80 to 100% of the load capacity.
Hard Calibration Ladder Logic Example
Verify that the parameters have been setup for your weighing process in accordance with
the setup information provided in Chapter 4.
A Ladder Logic example on the Hardy Process Solutions Inc. web site explains how to
set the weigh process parameters. Depending on your application, this Ladder Logic
example may or may not meet your requirements.
The Hard Calibration Ladder Logic Example is located at the Hardy Process Solutions
Inc. Web Site. See
http://www.hardysolutions.com
. Click products > PLC Weighing
Modules > HI 1756-FC. Click on the Docs & Programs tab and scroll down to Programs
and Firmware. Select
HI 1756-FC Sample Programs
for the Ladder Logic example.