Hanwei Alc Smart AT8100 Operation Manual Download Page 7

AT8100 Alcohol Tester operation manual 

6.4 Pass 

Input correct password in the password input mode, then press

button to enter into the 

“Pass  Value”  setup  interface.  Press 

  or 

button  to  adjust  the  pass  value,  and 

press

  button to save it. 

6.5 Fail 

The setup method is the same as “6.4 Pass”. 

6.6 Operator 

Input correct password in the password input mode, then press 

button to enter into the 

interface of register personnel number, the specific input method is the same as the above 

“password input method”. After the input is over, press

button to save it. 

6.7 Password 

Input accurately the former password in the password input mode, then press

button to 

enter  into  New  Password  input  interface.  After  input  the  new  password  as  the  above 

“Password  input  method”,  press

button  to  save  it.  Please  clearly  remember  the  new 

password after modification. 

6.8 Delete 

Input  correct  password  in  the  password  input  mode,  press

  button  and  the  screen 

indicates “Delete all

”, meanwhile, press

button to delete all the history records, press

C

button to give up the deleting. The screen will indicate “No records” and return to the 

main interface without history record. 

6.9 Tem. Scale 

This tester possesses “

”and“

”, the function of two temperature unit switch. Press 

 

or 

 

button  to  switch  the  temperature  unit  after  entering  the  “Tem.  Scale”  interface. 

When  switching  the  appropriate  unit,  press 

button  to  save  it,  and  the  temperature 

value in the main interface will transfer to the new unit at this moment. 

6.10 Next Cal

Enter into this option to look over the date of calibration last time and the days for the next 
calibration. 

7. Record upload 

This  function  is  able  to  upload  the  test  data  to  the  computer  for  preparation  management 
with software. The upload step as follows: 

Notice: 

As 

“Delete” operation will delete all the history records entirely and can not 

revert, please be cautious to operate.

 

Summary of Contents for Alc Smart AT8100

Page 1: ...169070 80 Fax 86 0371 67169090 Service E mail sales hwsensor com Website http www hwsensor com AT8100 HENAN HANWEI ELECTRONICS CO LTD Address No 169 Xuesong Rd National High tech Zone Zhengzhou 450001 China Service Tel 86 0371 67169070 80 Fax 86 0371 67169090 Service E mail sales hwsensor com Website http www hwsensor com HW120401146 ...

Page 2: ... POWER OFF 3 4 2 1 manual power off 3 4 2 2 power saving off 3 4 2 3 low battery power off 3 4 3 TEST 3 4 4 PRINT 3 5 TEST RECORD 4 6 SETUP 4 6 1 TIME 4 6 2 UNIT 4 6 3 PRINT 4 6 4 PASS 5 6 5 FAIL 5 6 6 OPERATOR 5 6 7 PASSWORD 5 6 8 DELETE 5 6 9 TEM SCALE 5 6 10 NEXT CAL 5 7 RECORD UPLOAD 5 8 BATTERY CHANGE 6 9 ORDINARY FAULT AND SOLUTION 6 10 PRECAUTION 6 11 ACCESSORIES 6 ...

Page 3: ...nced 16 bit low power consumption MCU Electrochemical sensor good reliability and accuracy Color LCD indication Passive test function Wireless printing function Real time and temperature indication Two level audible and visible alarm indication Low battery automatic power off function Blowing interruption and testing rejection function Unit and alarming level adjustable History record checking 200...

Page 4: ...mension l b h 122 67 31mm Weight about 120g excluding battery Table 1 Allowable error Alcohol gas concentration range C mg L Error C 0 400 0 032 mg L 0 400 C 1 000 8 C 1 000 30 Table 1 the allowable error in the process of detecting 4 Operation instruction 4 1 Power on Press the button the tester is powered on and it will make the self check automatically with a buzzer ringing the red and green in...

Page 5: ...e influenced After the analysis the screen displays concentration value and the follow several test types 1 Normal After testing finished if the testing result is under the pass level concentration value will indicate Clean 2 Pass After testing finished if the testing result reaches the pass level concentration value will indicate Pass and buzzing at the same time 3 Fail After testing finished if ...

Page 6: ...Unit Tem Scale Next Cal Input method of password press button to regulate the size of value in the password input mode change the numbers by press it every time till indicating the number you need press button to start the next number input During the process press button to cancel the number you choose press button to enter into option operational interface The original password of the tester is ...

Page 7: ...ification 6 8 Delete Input correct password in the password input mode press button and the screen indicates Delete all meanwhile press button to delete all the history records press C button to give up the deleting The screen will indicate No records and return to the main interface without history record 6 9 Tem Scale This tester possesses and the function of two temperature unit switch Press or...

Page 8: ...be used in specified conditions Long time no calibration Calibrate the tester Wrong Time Display Wrong time setup Reset the time Electromagnet misoperation No calibration for pressure sensors Calibrate pressure sensors again 10 Precaution Please read the manual carefully and follow the instruction before your testing To ensure the real testing result please wait 15 min to take the testing after dr...