Hanwei Alc Smart AT8100 Operation Manual Download Page 5

AT8100 Alcohol Tester operation manual 

4.2 Power off 

4.2.1 Manual power off 

 

In the power on mode, press 

button at least 1s, the tester will be powered off. 

 

Choose 

E

E

E

x

xx

i

ii

t

tt  with 

M

M

M

e

ee

n

nn

u

uu  interface, then press “

 

” button to confirm power off. 

4.2.2 Power saving off 
The  tester  will  be  powered  off  if  without  any  operation  for  more  than  3  minutes  at  main 
interface. 
4.2.3 Low battery power off 
The  tester  will  be  powered  off  if  the  battery  is  too  low,  please  take  out  the  battery  and 
recharge or store them. 

Indication of battery icon 

 

Full battery capacity 

 

Low battery capacity 

 

Partial battery capacity 

 

Short of battery capacity, recharge it timely 

4.3 Test 

As indicated in picture 1, put on a clean one-off mouthpiece. As picture 2, press

button 

in  the  main  interface,  the  screen  will  show  Waiting,  take  a  deep  breath  till  the  screen 
indicates “Blow Now”, blow equaly to the mouthpiece in the uniform continuity way till when 
hearing a ring. 
The  screen  will  display  “Processing”  when  blowing  finished,  remind  that  the  tester  is 
analyzing the breath gas sample. Do not blow at that moment, otherwise the result will be 
influenced. After the analysis, the screen displays concentration value and the follow several 
test types: 

1

) 

Normal:  After  testing  finished,  if  the  testing  result  is  under  the  pass  level, 
concentration value will indicate “Clean”. 

2

) 

Pass:  After  testing  finished,  if  the  testing  result  reaches  the  pass  level, 
concentration value will indicate “Pass”, and buzzing at the same time. 

3

) 

Fail: After testing finished, if the testing result reaches the fail level, concentration 
value will indicate “Fail”, and buzzing at the same time. 

4

) 

Passive 

test: If tester didn’t fit test or have ability to test, being taken a passive test. 

When the screen indicates “Blow Now”, take the mouthpiece off and the hole of 

blowing 

aim  to  the  tester’s  nostril.  After  2s,  press 

  button  to  enter  into 

passive test status. The screen indicates “Passive” and the result after analysis. 

5

) 

Test rejection: the tester wil

l buzz and indicate “Rejected

” without blowing after 

indicating “Blow Now” for 30s, meanwhile, press 

C

 

button to return to “Blow 

Now”  mode,  press

button  enter  into  display  for  result,  and  indicate 

“Rejected”. 

In  the  display  of  result  mode,  Press

button  to  save  result  and  enter  into  printing 

interface, press 

C

  button to cancel saving and return to the main interface. 

4.4 Print 

Press

button to enter into the printing interface in the display of result mode. The screen 

Summary of Contents for Alc Smart AT8100

Page 1: ...169070 80 Fax 86 0371 67169090 Service E mail sales hwsensor com Website http www hwsensor com AT8100 HENAN HANWEI ELECTRONICS CO LTD Address No 169 Xuesong Rd National High tech Zone Zhengzhou 450001 China Service Tel 86 0371 67169070 80 Fax 86 0371 67169090 Service E mail sales hwsensor com Website http www hwsensor com HW120401146 ...

Page 2: ... POWER OFF 3 4 2 1 manual power off 3 4 2 2 power saving off 3 4 2 3 low battery power off 3 4 3 TEST 3 4 4 PRINT 3 5 TEST RECORD 4 6 SETUP 4 6 1 TIME 4 6 2 UNIT 4 6 3 PRINT 4 6 4 PASS 5 6 5 FAIL 5 6 6 OPERATOR 5 6 7 PASSWORD 5 6 8 DELETE 5 6 9 TEM SCALE 5 6 10 NEXT CAL 5 7 RECORD UPLOAD 5 8 BATTERY CHANGE 6 9 ORDINARY FAULT AND SOLUTION 6 10 PRECAUTION 6 11 ACCESSORIES 6 ...

Page 3: ...nced 16 bit low power consumption MCU Electrochemical sensor good reliability and accuracy Color LCD indication Passive test function Wireless printing function Real time and temperature indication Two level audible and visible alarm indication Low battery automatic power off function Blowing interruption and testing rejection function Unit and alarming level adjustable History record checking 200...

Page 4: ...mension l b h 122 67 31mm Weight about 120g excluding battery Table 1 Allowable error Alcohol gas concentration range C mg L Error C 0 400 0 032 mg L 0 400 C 1 000 8 C 1 000 30 Table 1 the allowable error in the process of detecting 4 Operation instruction 4 1 Power on Press the button the tester is powered on and it will make the self check automatically with a buzzer ringing the red and green in...

Page 5: ...e influenced After the analysis the screen displays concentration value and the follow several test types 1 Normal After testing finished if the testing result is under the pass level concentration value will indicate Clean 2 Pass After testing finished if the testing result reaches the pass level concentration value will indicate Pass and buzzing at the same time 3 Fail After testing finished if ...

Page 6: ...Unit Tem Scale Next Cal Input method of password press button to regulate the size of value in the password input mode change the numbers by press it every time till indicating the number you need press button to start the next number input During the process press button to cancel the number you choose press button to enter into option operational interface The original password of the tester is ...

Page 7: ...ification 6 8 Delete Input correct password in the password input mode press button and the screen indicates Delete all meanwhile press button to delete all the history records press C button to give up the deleting The screen will indicate No records and return to the main interface without history record 6 9 Tem Scale This tester possesses and the function of two temperature unit switch Press or...

Page 8: ...be used in specified conditions Long time no calibration Calibrate the tester Wrong Time Display Wrong time setup Reset the time Electromagnet misoperation No calibration for pressure sensors Calibrate pressure sensors again 10 Precaution Please read the manual carefully and follow the instruction before your testing To ensure the real testing result please wait 15 min to take the testing after dr...

Reviews: