background image

CC-800 AC/DC CURRENT PROBE 

OPERATOR’S

 

MANUAL 

 

Summary of Contents for CC-800

Page 1: ...CC 800 AC DC CURRENT PROBE OPERATOR S MANUAL ...

Page 2: ... is powered normally the green power indicator will light up 3 When the measured current is not greater than 20 amperes please set the current probe to the 10mV A range and select the AC 200mV or DC 200mV range on the multimeter according to the type of current being measured 4 When measuring current it is necessary to perform zero adjustment before measuring That is to say rotate the zero adjustm...

Page 3: ... to the negative terminal of the current probe 2 When performing DC current measurement the current probe may not reach the correct zero point due to the hysteresis effect In order to eliminate this effect you need to open and close the jaws of the current probe normally several times and then adjust the zero adjustment knob to approach the zero point APPLICATION SAFETY 1 The voltage of the conduc...

Page 4: ...f its accuracy is 0 5 the maximum value displayed by the digital multimeter is 10 20mV and the minimum is 9 80mV 10 15mV 1 0 5 10 20mV 9 85mV 1 0 5 9 80mV Accuracy of current probe DC current 10mV A 1 5 0 5A 1A 40A DC current 1mV A 1 5 0 5 A 1A to 800 A AC current 0Hz to 500Hz 10mV A 2 0 5A 1A 40A AC current 0Hz to 5KHz 1mV A 2 0 5A 1A 500A Load resistance 10KΩ typical Temperature coefficient 0 1 ...

Page 5: ...The products comply with the following EU standards 89 336 EEC electromagnetic compatibility and 93 68 EEC CE marking improved 73 23 EEC low voltage 5 ...

Reviews: