INTRODUCTION

 

The multimeter/oscilloscope can directly measure the DC current of no more than 800A 

and the AC current of no more than 500A through the AC/DC current probe, and the 

measurement frequency is up to 5KHz. When measuring current with this current probe, 

there is no need to destroy the current loop or peel off the insulating layer. 

The current probe is not only suitable for the detection or monitoring of possible leakage in 

the circuit, but also the extended jaw can be measured in a small gap. It should be noted 

that when measuring DC current, it is necessary to adjust the zero point by operating the 

zero adjustment knob. 

APPLICATION PROCEDURE 

1.  When connecting with a multimeter, insert the GND terminal of the BNC/banana-KJJ 

adapter into the COM terminal of the multimeter, and insert the other terminal of the 

adapter into the V-

Ω terminal. Please note that the input impedance of the multimeter 

used should be at least 10K ohms. 

When connecting with the oscilloscope, connect the BNC plug of the current probe to 

the BNC terminal of the oscilloscope. 

2.  Set the power switch from the "OFF" to the range required to measure the current, 

that is  to say, push the switch slider to 10mV/A  or  1mV/A range. When the current 

probe is powered normally, the green power indicator will light up. 

3.  When the measured current is not greater than 20 amperes, please set the current 

probe to the 10mV/A range, and select the AC 200mV or DC 200mV range on the 

multimeter according to the type of current being measured. 

4.  When measuring current, it is necessary to perform zero adjustment before 

measuring. That is to say, rotate the zero adjustment knob of the current probe until 

the reading on the multimeter is close to zero. When connecting an oscilloscope to 

measure DC current, set DC coupling. Before measuring, rotate the zero adjustment 

button of the current probe until the oscilloscope's glow line returns to the reference 

position, and then release the zero adjustment button. 

5.  When the 10mV/A range  of the current probe is selected, the data measured by a 

multimeter or oscilloscope (in mV) multiplied by 100 is the actual current value, and 

the unit of the current value is mA. For example, if the value measured by the current 

multimeter is 10mV, the actual current value of the current-carrying conductor under 

test is 10×100=1000mA. 

When the 1mV/A range  of the current probe is selected, the data measured by a 

multimeter or oscilloscope is multiplied by 1000 to obtain the actual current value, and 

the unit of the current value is mA. For example, if the value measured by the current 

multimeter is 5mV, the actual current value of the current-carrying conductor under 

 

Summary of Contents for CC-800

Page 1: ...CC 800 AC DC CURRENT PROBE OPERATOR S MANUAL ...

Page 2: ... is powered normally the green power indicator will light up 3 When the measured current is not greater than 20 amperes please set the current probe to the 10mV A range and select the AC 200mV or DC 200mV range on the multimeter according to the type of current being measured 4 When measuring current it is necessary to perform zero adjustment before measuring That is to say rotate the zero adjustm...

Page 3: ... to the negative terminal of the current probe 2 When performing DC current measurement the current probe may not reach the correct zero point due to the hysteresis effect In order to eliminate this effect you need to open and close the jaws of the current probe normally several times and then adjust the zero adjustment knob to approach the zero point APPLICATION SAFETY 1 The voltage of the conduc...

Page 4: ...f its accuracy is 0 5 the maximum value displayed by the digital multimeter is 10 20mV and the minimum is 9 80mV 10 15mV 1 0 5 10 20mV 9 85mV 1 0 5 9 80mV Accuracy of current probe DC current 10mV A 1 5 0 5A 1A 40A DC current 1mV A 1 5 0 5 A 1A to 800 A AC current 0Hz to 500Hz 10mV A 2 0 5A 1A 40A AC current 0Hz to 5KHz 1mV A 2 0 5A 1A 500A Load resistance 10KΩ typical Temperature coefficient 0 1 ...

Page 5: ...The products comply with the following EU standards 89 336 EEC electromagnetic compatibility and 93 68 EEC CE marking improved 73 23 EEC low voltage 5 ...

Reviews: