Attach the standard xReceiver to the metal adapter plate�
How to measure performance in a coupler.
RIC devices should be measured with a HA1 coupler since it best reflects the
acoustical properties in the ear canal� The xReceiver kit includes a metal adapter
plate that fits on a HA1 coupler and allows for a good connection of the
xReceiver unit to the coupler�
Select Activate Test Box Measurement Mode in HANSATON scout software
from the menu on the top menu bar� Clicking on Activate Test Box Measure-
ment Mode disables the adaptive features and activates the omnidirectional
microphone mode to allow for correct measurements�
Click on Deactivate Test Box Measurement Mode to reestablish the
previous settings�
18
Fitting guide.
Receiver-in-Canal (RIC).