63
62
• To set the sample number (choice #1),
press 1 followed by ENTER. "
P1.1
" will
be displayed on the secondary LCD with
"
0
" on the primary one.
Enter the desired number (choose any
number from 1 to 999) followed by EN-
TER. E.g. sample number 200.
The printer will print the value.
Samples after 999 will be reset to zero
and the sample number will automatically
advance from 0 to 999 in increment of 1
each time the instrument logs and/or prints
a reading.
PROGRAM 1
1 set sample no.
2 set time
3 set date
1
ENTER/
2
0
0
ENTER/
PROGRAM 1
1 set sample no.
2 set time
3 set date
Sample number
200
Program #7
To select a different printing language (6
languages are available) (see page 78).
Program #8
To print the ion concentration in M for the
measured pH value (see page 79).
Program #9
To set the baud rate and the STRINGII code
of Escape character for the RS232 (see
page 79).
To stop or exit any program:
•
press PGM
•
press 0
Note: the instrument must be in pH mode for
program #2, #3 and #8. Otherwise a
long beep will sound.
PROGRAM #1 (pH 302)
This program allows you to set sample num-
ber, time and date.
To enter program #1:
• Push the PGM key and then 1.
The printer will then define the three func-
tions:
1 is the sample number,
2 is time,
3 is the date.
PGM
0
1
PGM