19
OUR Configuration
The
OUR
method configuration allows the user to set the parameters related to the
OUR
measurement.
OUR
is used to determine the oxygen consumption or respiration rate. It is defined as the mg/L of
oxygen consumed per hour.
These parameters will be used to flag error limits and make dilution calculations for the determination.
Ignore this parameter if not making
OUR
measurements.
Highlight
OUR configuration
and press
Select
.
SETUP
SETUP
Press / keys to modify the selected parameter’s value.
Press
Prev/Next
to select a different parameter.
Press
Save
to save the new
BOD
configuration.
Press
ESC
to leave without changing.
Parameters:
•
Sample min
∆
DO
‑ the minimum acceptable difference between the initial and final
DO
values
for a sample. If the difference is less than this value the meter will show a warning message when
evaluating the
BOD
.
Range: 0.00 to 50.00 mg/L.
•
Sample min end DO
‑ the minimum acceptable final
DO
value for a sample. If the final
DO
value
is less than this value the meter will show a warning message when evaluating the
BOD
.
Range: 0.00 to 50.00 mg/L.
•
Seed min
∆
DO
‑ the minimum acceptable difference between the initial and final
DO
values
for a seed sample. If the difference is less than this value the meter will show a warning message
when evaluating the
BOD
.
Range: 0.00 to 50.00 mg/L.
•
Seed min end DO
‑ the minimum acceptable final
DO
value is less than this value the meter will
show a warning message when evaluating the
BOD
.
Range: 0.00 to 50.00 mg/L.