46
Temperature Source
Highlight
Temperature Source
.
Press the displayed functional key in order to change the option.
Select
Probe
in order to take the temperature automatically with the temperature sensor inside the
electrode.
Select
Manual
in order to set the temperature using the
/
keys.
Temperature Compensation
Highlight
T. Compensation
.
Press one of the selected functional keys to change the option.
Select
No TC
in order to display actual conductivity (no temperature compensation).
Select
Linear
in order to automatically compensate conductivity using the set temperature coefficient.
Select
Non Linear
temperature compensation for natural water measurements, using the natural
water compensation equation.
Range Select
Highlight
Range Select
.
SETUP