26
Press
Continue
to enter
USP Stage 3
analysis.
The instrument will display the sample pH setting mode.
Use a calibrated pH meter to read the pH value of the sample.
Use / keys to set the value to that displayed on the pH meter.
Press
Accept
to confirm the pH setting.
The
USP Stage 3
report will be displayed.
The report will include all stages information.
Press
key to scroll the report pages .
Press
Log
to store the report.
Press
ESC
to return to
USP
main screen.
Notes: The Stage 3 report will include the information about Stage 1 and Stage 2.
If the log space is full enter view logged data mode by pressing RCL key and free log
space by deleting previously stored records.
Press
Stage 3
to enter in Stage 3 water analysis.
The
USP Stage 3
tutorial is displayed.
USP MODE PROCEDURES
Press
Log
to store report.