ELECTRODE MAINTENANCE
11. ELECTRODE MAINTENANCE
Code and serial number
on the other side
Temperature sensor
Quick connector with
rubber protection and 1 meter cable
Reference junction
pH bulb
11.1. PREPARATION PROCEDURE
1. Remove the protective cap. Do not be alarmed if salt deposits are present, this is normal. Rinse
the electrode with water.
2. Shake the electrode down as you would do with a clinical thermometer to eliminate any air
bubbles inside the glass bulb.
3. Remove the fill hole cover to ensure the reference junctions are flowing. Set aside for storage.
4. Top off the electrolyte filling solution using Hanna
HI7082
Electrolyte solution.
5. If the glass bulb and/or junction are dry, soak the electrode in
HI70300
Storage solution for a
minimum of 30 minutes.
6. Rinse with water and calibrate before using.