17
17
SETUP
Press
Add
to add a custom buffer to the list (maximum five).
Use the / keys to change the custom buffer value. Press
Accept
to confirm the value or press
ESC
key to leave without saving.
If custom buffers have been added previously, use the / keys to select a calibration buffer.
Press
Modify
to edit the buffer value.
Press
Delete
to delete the selected custom buffer from the list.
Press
Add
to add buffer to the list (maximum five).
View Calibration Points
Option: Enabled or Disabled
(default option: Enabled)
Press the functional key to change to desired option.
If enabled, the buffers used in the last calibration are displayed on the pH measurement screen.
Out of Calibration Range
Option: Enabled or Disabled
(default option: Enabled)
Press the functional key to change to desired option.
If enabled, when the pH reading is outside the calibration range, the “Out of calibration range”
message is displayed.