33
Out of Calibration Range Warning
Highlight
Out of Cal.Range Warning
.
Press the displayed functional key in order to change option.
If enabled, the
“Out Cal Range”
message will be displayed if the pH reading is outside. Approxi‑
mately 1 pH unit, from the lowest and highest pH buffer value used for calibration. For single point
pH 7.01 calibration the message will be displayed for pH values less than 4 or greater than 10.
SETUP